5. Programming Procedure
5.2.5
Process Step: END
C:Y P: X END
- + other yes
Cycle Y END ?
no yes
Nr. of Cycles: 1
- + yes
Different Cycle?
no yes
Final Aspirate ?
no yes
Time: 4 sec
- + exit yes
Speed: 10mm/s
- + exit yes
Crosswise: NO
- + exit yes
Print Prot.? ASK
- + exit yes
C:Z P:1 END
- + other yes
Select number of cycles to be performed
using these programmed steps (1 - 9)
Select if a different cycle
is to be defined
Select if the microplate is to be aspirated
at the end of the procedure
Select required aspirating time (1 - 20 sec)
Select and confirm End step
Confirm end of programming for this cycle
Select required aspirating speed (1 - 20 mm/s)
Select if crosswise aspiration is to be used
If printer option fitted, select setting for
protocol printing (YES, NO, ASK)
Select steps for next cycle
Run X:yyyyyyy
- + other yes
Instrument ready to perform defined
washing procedure
Flat bottom
wells
Round bottom
wells
May 2003
Instructions for Use for Columbus Washer No. I 109 006 Rev No. 2.2
5-5