4. Operating Instructions
4.6 End of Operation
At the end of operation, the instrument can be either left switched on with the
manifold in the rinsing solution or switched off.
Instrument Left Switched On
Use the following procedure to leave the instrument switched on with the
manifold soaking in the distilled water at the end of operation:
Place the wash solution tube into a bottle containing distilled water.
Use the
Rinse: Night
rinsing procedure to rinse the wash system.
See chapter 7.
When the rinsing procedure is completed the following message is displayed:
Soaking
Stop
If the manifold is to be soaked overnight, do not switch the instrument off.
Instrument Switched Off
If the instrument is to be switched off at the end of operation the
Rinse: Day
rinsing procedure should be performed to prevent the needles becoming blocked.
Place the wash solution tube into a bottle containing distilled water.
Use the
Rinse: Day
rinsing procedure to rinse the wash system.
When the rinsing procedure is completed, switch the instrument off.
4-14
Instructions for Use for Columbus Washer No. I 109 006 Rev No. 2.2