3 - 18
Cavro® XLP 6000 Modular Syringe Pump Operating Manual, 734237-C, en
3 - Software Communication
Communication Protocols
Multi-Frame Command
The host commands [Z2S5gIA3000OgHD300G10G5R] to a pump, and the pump
is set to address 0.
Note: For multi-frame commands, the pump only acknowledges the last frame.
Common Command
After the host has sent command [A1000A0] to the pump, it sends command 0 of
frame type 2 to a pump and makes the pump move. The pump is set to address 0.
Query Command
The host commands 29 of frame type 6 to a pump, and the pump is set to address
1.
Host sends:
0
XXXXXXX
011
0
1000
Z2S5gA3
Dir
Slave ID
Frame type
RTR
DLC
Data bytes
0
XXXXXXX
100
0
1000
000OgHD30
Dir
S
lave ID
Frame type
RTR
DLC
Data bytes
0
XXXXXXX
001
0
1000
00G10G5R
Dir
Slave ID
Frame type
RTR
DLC
Data bytes
Pump acknowledges:
1
XXXXXXX
001
0
0000
Dir
Slave ID
Frame type
RTR
DLC
After executing the command, pump reports status:
1
XXXXXXX
001
0
0010
<20h><60h>
Dir
Slave ID
Frame type
RTR
DLC
Data bytes
Host sends:
0
010
0000
010
0
0001
1
Dir
Group
Device
Frame type
RTR
DLC
Data bytes
Pump does not acknowledge common commands.