44
7. Comparison Setting
7-3. Hysteresis
This function applies a width to switching from ON to OFF for high limit and low limit judgment display.
If the indicator value fluctuates near a value, causing judgment display to chatter between ON and OFF, adjustment of the hysteresis
value can prevent this chattering.
The operation of hysteresis changes as follows according to the Comp. Pattern Setting.
For limit boundary values that are greater than the OK position, it operates in the negative direction. For limit boundary values that
are less than the OK position, it operates in the positive direction.
When the Comp. Pattern Setting is LO/OK/HI, hysteresis activates for the high limit value when an indicator value decreases.
OFF
ON
OFF
ON
OFF
ON
OFF
ON
Indicator
value
Indicator
value
Time
Time
Judgment
output
Judgment
output
High limit
Judgment output if
hysteresis inactive
Judgment output if
hysteresis active
High limit
Low limit
High limit
Hysteresis
Hysteresis
HI
HI
LO
Judgment display conditions (when Comp. Output Pattern is set to Standard Output)
Judgment
type
Status
Condition
HI
OFF
w
ON
High limit value < indicator value
ON
w
OFF
Indicator value ≤ (high limit value − hysteresis setting value)
LO
OFF
w
ON
Indicator value < low limit value
ON
w
OFF
(Low limit value + hysteresis setting value) ≤ indicator value
ATTENTION
Set the hysteresis setting value so that it meets the following conditions.
Low limit value < (high limit value − hysteresis setting value)
NOTE
o
The same Hysteresis Width setting is used with high limit and low limit values.
o
If set to "0", hysteresis will be disabled.