HANDY CALIBRARTOR   

CA-20 

 

 

Instruction Manual 

 

 

Please read this manual before using the product 

and keep the manual handy.

 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Rev. D 

 

 

 

 
 

TEAC Corporation   February, 2012

 

Содержание CA-20

Страница 1: ...HANDY CALIBRARTOR CA 20 Instruction Manual Please read this manual before using the product and keep the manual handy Rev D TEAC Corporation February 2012 ...

Страница 2: ...ilt in installation such as a bookcase or rack unless proper ventilation is provided or the manufacturer s instructions have been adhered to 10 Power Sources This product should be operated only from the type of power source indicated on the marking label If you are not sure of the type of power supply to your home consult your product dealer or local power company For products intended to operate...

Страница 3: ...ed technician to restore the product to its normal operation e if the product has been dropped or damaged in any way f when the product exhibits a distinct change in performance this indicates a need for service 18 Replacement Parts When replacement parts are required be sure the service technician has used replacement parts specified by the manufacturer or have the same characteristics as the ori...

Страница 4: ...on TEAC CORPORATION disclaims any warranty that any use of such data and or information shall be free from infringement of any third party s intellectual property rights or other proprietary rights and further assumes no liability of whatsoever nature in the event of any such infringement or arising from or connected with or related to the use of such data and or information Microsoft Windows Wind...

Страница 5: ...rmits such activity You may not rent or lease the SOFTWARE You may not reproduce the SOFTWARE except for archival purpose 4 TERMINATION Your rights under this Agreement terminate upon the disposal of all copies of the SOFTWARE or without prejudice to any other rights TEAC may terminate this Agreement if you fail to comply with the terms and conditions of this Agreement In such event you must destr...

Страница 6: ...4 Operations 6 4 1 Install batteries 6 4 2 Attach accelerometer 6 4 3 Driving 7 4 4 Measure 8 5 Function 10 5 1 Storing Data to CA 20 10 5 2 Data List 10 5 3 Measurement Mode 10 5 4 Auto Power Off 10 5 5 Time Setting 10 5 6 Storing Data to PC 11 6 Specification 11 7 Dimension 12 2 ...

Страница 7: ...rs The unit shakes the accelerometer measures output signal from accelerometer and calculates sensitivity CA 20 shakes accelerometer at 159 2 Hz to get sensitivity easily and simply in the field It is not recommended for high precision calibration for laboratory or institute use 2 Names and Parts Rear Side 3 ...

Страница 8: ...ty data to PC or CA 20 On menu mode this works as ENTER key 5 PC Link Red Communication mode is set to Link Blue Communicating with PC Off Communication mode is set to Cut 6 LCD Display Sensitivity bias voltage mode etc are displayed 7 Shaker Base which accelerometer is connected to 8 Signal input connector BNC Output signal from accelerometer has to be connected to 9 DC input connector Connect DC...

Страница 9: ...ist view 5 Clear Data List Clear Cancel Clear all data 6 Measure Mode Single Repeat Select measure mode 7 Auto Power off ON 3min OFF Set auto power off 8 LCD illumination ON OFF Select LCD backlight setting 9 Date Time Set Time Date Set Time Date Month Year 10 Data Transmit Transmit Cancel Transfer data to PC Return Return Factory settings All settings are kept even if CA 20 is turned off Typical ...

Страница 10: ...lkaline batteries Use AC DC adapter when it s used for a long time Remove batteries if it s not used for a while 4 2 Attach accelerometer Clean the base and remove dust If a accelerometer has a female screw attach accelerometer using screw which supplied with accelerometer If a accelerometer has M6 male screw it can be attached using male screw If a screw which accelerometer has is not M6 use conv...

Страница 11: ...erometer push select switch to change charge mode In case of IEPE ICP type accelerometer push select switch Vto change ICP mode And select power supply such as voltage and current for accelerometer Note Make sure power supply such as voltage and current for accelerometer If these are too high accelerometer might get damaged 7 ...

Страница 12: ...m Overload G G 1 Sensitivity Calculated sensitivity is displayed 2 Unit pC G Charge mode mV G IEPE ICP mode 3 Accelerometer type Charge Voltage ICP Charge Charge mode Voltage IEPE ICP mode 4 Supply voltage current ICP supply voltage current 8 ...

Страница 13: ...te If this alarm appears calibrator stops to shake 7 Low battery alarm Installed batteries voltage are low change batteries or connect AC DC adapter 8 Over range Sensitivity is too high maximum sensitivity which CA 20 can calibrate is 196 pC G or 196 mV G Minimum sensitivity which CA 20 can calibrate is 0 255 pC G or 0 255 mV G 9 ...

Страница 14: ...tion 5 3 Measurement Mode Two measurement mode Single and Repeat are available Single means single shot measurement After measured sensitivity being stable CA 20 stops to shake base Push select switch bit longer to measure sensitivity again Repeat means continuous measurement CA 20 doesn t stop to measure sensitivity Use Repeat mode in case of measuring low sensitivity accelerometer 5 4 Auto Power...

Страница 15: ...e depth of 6 5 mm Operating temperature range 10ºC to 40ºC Storage temperature range 10ºC to 50ºC Operating Altitude Max 2000 meters Batteries LR 6 x 4 batteries 5V DC 0 75A Continuous use approx 1 hours Calibration sensor weight approx 25g Battery life LCD ON Power supply for accelerometer Selectable DC 24 15 V 0 5 2 4mA Constant current supply AC Adapter AC100 240V 50 60Hz 0 3A Power supply for ...

Страница 16: ...7 Dimension 12 ...

Страница 17: ...Copyright 2012 TEAC Corporation All rights reserved TEAC CORPORATION 1 47 Ochiai Tama shi Tokyo 206 8530 Japan E mail tic_cs teac co jp 13 ...