background image

408-32113

2

 of 2

Rev 

A

CAUTION

Never insert the contact upside down; otherwise, the 

locking latches could depress and lodge into the cavity 

in an unseated position or become compressed so that 

they will not lock when repositioned for proper insertion.

b. Push the contact straight, not rotated, into the 
contact cavity. If it is rotated, damage to the 
contact will occur or the contact will not latch 
properly into the contact cavity. See Figure 2B. 
Do not force the contact into the housing. It 
should slide in and latch into position with 
minimal force.

c. Ensure that the contact latches are locked into 
the cavity by pulling lightly on the wire and 
checking for resistance to pull out.

3. Install the TPAs into the housing as follows:

a. Align a TPA with the wire end of the housing 
so that each tooth faces a contact cavity with its 
radius surface against the wire. See Figure 3A.

b. Slide the TPA into the housing so that the 
teeth enter the contact cavities evenly and until 
all the locking latches lock onto the ledges 
around the perimeter of the housing. Make sure 
to push on each locking latch individually to 
ensure complete installation. There will be an 
audible and tactile “click”. See Figure 3B.

NOTE

TPAs with more than 5 positions have 4 locking latches. 

TPAs with 5 positions or less have only 2 locking 

latches. Make sure to push on each locking latch 

individually to ensure complete installation.

c. Flip the housing over and repeat steps “3a” 
and “3b” with a second TPA for the other row.

4. EXTRACTING CONTACTS

1. TPAs can be removed from the housing by 
releasing the locking latches from the ledges and 
pulling the TPA straight away from the housing. No 
contact can be extracted until the TPA is removed.

2. Contacts can be removed from housings using 
Extraction Tool 1586343-1 according to Instruction 
Sheet 408-32114.

CAUTION

DO NOT re-insert any extracted contacts.

5. REPLACEMENT AND REPAIR

DO NOT use defective or damaged products. 

Contacts, housings, and TPAs are not repairable.

DO NOT re-use a contact by removing the wire.

DO NOT re-insert any contact after extracting it 

from the receptacle housing.

6. REVISION SUMMARY

Initial release of document

Figure 3

!

NOTE

i

!

3A

3B

Align TPA with Teeth 

Facing Contact 

Cavities

Place Radiused 

Surfaces of Teeth 

Against the Wires

Ledge Around 

Perimeter of Housing

Latch Fully 

Engaged 

Onto Ledge

Отзывы: