RBK-X1C
PN 2376800-1
Page 34 Rev. A Dec. 2020
4.2.8
Auto Centering Operation
Action of the Process Start buttons is inhibited until the heater LED has
changed to green.
Operation of any individual Start button during the timer count-down will
manually override the process and move the heater chamber to the rear.
Fig. 4-3: Auto Centering
1. Log on in HMI, switch ON to the
‘Centering’
button in Maintenance page, to activate Auto
Centering function (see section 3.5, item 4).
2. Turn back to main interface, you would find it show additional auto centering illuminator on the
screen top (see section 1.3, item 13). Press both start button to drop down the DETECTION
PLATE, the centering illuminator will show green.
3. Pull and rotate the right DISTANCE KNOB to find a suitable distance between 2 detection plates,
according to the width of splice, then processor is ready for centering.
4. If necessary, pull and rotate the left OFFSET KNOB to adjust the splice offset.
Note: Due to current centering device capacity, splice offset CANNOT lager than 10mm.
5. Push tube to expose the splice (Fig.4-3), then insert cable with splice contact with both
DETECTION PLATE.
Note: Due to current centering device capacity, tube CANNOT longer than 60mm.
6. When detection success, the DETECTION PLATE will swing up, then the CENTERING PLATE
will push the tube to the center, and back to both side at once.
Note: In case of CENTERING PLATE was stuck when moving to center, remove the cable manually and
press on both start button to start a new cycle.
7.
Press on both start buttons simultaneously 0.5S, heater carrier will move forward to load position.
Note: If turn on
“centring
-
AUTO”
icon in maintenance page
, operator don’t need to press both start
button to activate heater carrier moving forward, it would automatically move forward when detection
success.
But we won’t recommend operat
ing like this, since heater carrier moving without hands on
button, operator may have chance to put hands in heating chamber. That would cause accident.
8.
Check from main interface that the process Cycle indicator lit green. Timer Display start to count
down.
Detection plate
Distance Knob
Offset Knob
Содержание RBK-X1C
Страница 55: ...RBK X1C PN 2376800 1 Page 54 Rev A Dec 2020 6 Schematic Diagram...
Страница 56: ...RBK X1C PN 2376800 1 Page 55 Rev A Dec 2020...
Страница 57: ...RBK X1C PN 2376800 1 Page 56 Rev A Dec 2020...
Страница 58: ...RBK X1C PN 2376800 1 Page 57 Rev A Dec 2020...
Страница 59: ...RBK X1C PN 2376800 1 Page 58 Rev A Dec 2020...
Страница 60: ...RBK X1C PN 2376800 1 Page 59 Rev A Dec 2020...
Страница 61: ...RBK X1C PN 2376800 1 Page 60 Rev A Dec 2020...
Страница 62: ...RBK X1C PN 2376800 1 Page 61 Rev A Dec 2020...
Страница 63: ...RBK X1C PN 2376800 1 Page 62 Rev A Dec 2020...
Страница 64: ...RBK X1C PN 2376800 1 Page 63 Rev A Dec 2020 7 Pneumatic Diagram...