background image

3-38

System 3

RX7 Stimulator Base Station

The front panel VFD screen reports detailed information about the status of the 

system. The display includes two lines. The top line reports the system mode, Run! 

or Idle, and displays heading labels for the second line. The second line reports the 

user’s choice of status indicators for each DSP followed by an aggregate value. 
The user can cycle through the various status indicators using the Mode button to 

the left of the display. Push and release the button to change the display or push 

and hold the button for one second then release to automatically cycle through each 

of the display options. The VFD screen may also report system status such as 

booting status (Booting DSP) or alert the user when the device's microcode needs 

to be reprogrammed (Firmware Blank).

Status

 

Indicators

Cyc:

 cycle 

usage

Ovr:

processor cycle overages

Bus%:

percentage of internal device's bus capacity used 

I/O%:

percentage of data transfer capacity used 

Important!

The status lights will flash (~3 times a second) to alert the user when a device 

goes over the cycle usage limit, even if only for a particular cycle. This helps to 

identify periodic overages caused by components in time slices.

Fiber

 

Optic

 

Output

 

Port

 

(Stimulator)

The output port, labeled Stimulator, can be used to transfer microstimulation 

waveforms to the MS16/MS4 Stimulus Isolator and/or to control its digital output. 

See the “MS4/MS16 Stimulus Isolator” on page 7-3.

Important!

This fiber optic port is disabled if the sampling rate of the system is set to a value 

greater than ~25 kHz. 

Fiber

 

Optic

 

Input

 

Ports

 

(Amp

A

 

and

 

Amp

B)

The RX7 base station can acquire digitized signals from a Medusa preamplifier over 

a fiber optic cable. This provides loss-less signal acquisition between the amplifier 

and the base station. Up to two fiber optic ports are provided to support 

simultaneous acquisition from up to two preamplifiers. Each port can input up to 16 

channels at a maximum sampling rate of ~25 kHz. The fiber optic ports provide 

oversampling. See “Fiber Oversampling”, below for more information. 
The fiber optic ports can be used with any of the Medusa preamplifiers including the 

RA16PA, RA4PA, or RA8GA. The channel numbers for each port begin at a fixed 

offset regardless of the number of channels available on the connected device. 

Channels are numbered as follows:

Amp-A

1 – 16

Amp-B

17 – 32

Fiber

 

Oversampling

 

(acquisition

 

only)

The fiber optic cable that carries the signals to the fiber optic input ports on the 

RX7 has a transfer rate limitation of 6.25 Mbits/s. With 16 channels of data and 16 

bits per sample, this limitation translates to a maximum sampling rate of ~25 kHz.

Содержание RX7

Страница 1: ...3 The RX7 includes 40 bits of digital I O analog output and can include one or two fiber optic input ports for acquisition of digitized data from Medusa preamplifiers Acquired signals can be filtered...

Страница 2: ...e information from the I O controller and the shared memory The master DSP supervises overall system boot up and operation The master DSP also acts as the main data interface between the zBus host PC...

Страница 3: ...delay is taken care of for the user in OpenEx when Timing and Data Saving macros are used See MultiProcessor Circuit Design in the RPvdsEx Manual for these and other multiprocessor circuit design tec...

Страница 4: ...Port Stimulator The output port labeled Stimulator can be used to transfer microstimulation waveforms to the MS16 MS4 Stimulus Isolator and or to control its digital output See the MS4 MS16 Stimulus I...

Страница 5: ...he connected amplifiers When an amplifier is not connected the Amp light will flash in a slow steady pattern The light is lit when the amplifier is connected and begins to flash quickly when the volta...

Страница 6: ...n upsampler for improved audio playback The upsampler is controlled through one of the RX7 s programmable bits and can be turned off to allow the D A to drive external devices such as a stimulator Cha...

Страница 7: ...x and select the RX7 Elec Stimulator from the list The dialog expands to display the Device Configuration Register 3 Click Modify to display the Edit I O Setup Control dialog box In this dialog box a...

Страница 8: ...four bytes of digital I O XLink The XLink is not supported at this time RX7 Technical Specifications The RX7 is designed for use with the MS16 stimulus isolator The RX7 is also equipped with a fiber o...

Страница 9: ...or Five Memory 128 MB SDRAM Shared D A 4 channels 16 bit PCM Sample Rate Up to 97 65625 kHz 8X upsampled to 200 kHz default operation Frequency Response DC Nyquist 1 2 sample rate Voltage Out 10 0 Vo...

Страница 10: ...ts 0 2 4 and 6 10 D1 Byte D Word addressable digital I O Bits 1 3 5 and 7 23 D2 11 D3 24 D4 12 D5 25 D6 13 D7 Pin Name Description Pin Name Description 1 BA0 Bit Addressable digital I O Bits 0 2 4 and...

Отзывы: