background image

6-8

System 3

PZ2 PreAmp

*Note: When sampling at a rate of 48.828 kHz the PZ2 preamplifier is limited to a 

maximum of 128 channels.
**Note: If longer cable lengths are required, contact TDT.

Input

 

Connectors

PZ2 Preamplifiers have up to 16, 26-pin headstage connectors on the back of the 

unit. A1 – A16 represent the 16 channels coming from each connected headstage. 

The PZ2 channels are marked next to the respective connector on the preamplifier. 

So, for the connector for channel 1 – 16, A1 is channel 1 while on the connector 

for channels 17 – 32, A1 is channel 17. 

Important!

Each input connector uses its own unique ground and reference. When using multiple 

headstages, ground pins on all headstages should be connected together to form a 

single common ground. See “Headstage Connection Guide” on page 6-99.

Distortion (typical)

< 1%

A/D Sample Rate

Up to 48828.125 Hz*

Sample Delay

Dependent on Sample Rate and RPvdsEx input method

Rate 

Pipe Input

MC Input

6 kHz

16 samples

15 samples

12 kHz

17 samples

16 samples

25 kHz

20 samples

19 samples

50 kHz

26 samples

25 samples

Input Impedance

10

5

 Ohms

Power Requirements

2 Lithium Ion cells at 10 AmpHours each

Battery

Eight hours to charge both cells

Battery life between charges, per cell:

32 ch ~ 13 hrs

64 ch ~ 11 hrs

96 ch ~ 9.5 hrs

128 ch ~ 8 hrs

256 ch ~ 5 hrs

Charger

External 6 V, 3 A power supply

Indicator LEDs

Up to 256 status or clip warning, battery life, active battery 

bank

Input referred noise 

μ

V rms typical 300- 7000 Hz, 8 

μ

V peak typical

Fiber Optic Cable

5 meters standard, cable lengths up to 20 meters**

Содержание PZ2

Страница 1: ...channel models Note When sampling at a rate of 50 kHz only the first 128 amplifier channels will be available System Hardware All PZ2 channels are organized into groups of 16 channel banks with each b...

Страница 2: ...hile collecting data Powering ON To turn the preamplifier on move the three position battery switch located on the front panel of the PZ2 to either the Bat A or Bat B position Powering OFF To turn the...

Страница 3: ...o indicate a clip warning Green Activity Red Clip Warning Clip Warning When the input to a channel is greater than 3dB from the preamplifier s maximum voltage input the LED for the corresponding chann...

Страница 4: ...eads to remove any potential RF noise that might travel through the cable For best results position the ferrite bead close to the source of the RF noise Battery Overview The PZ2 preamplifier features...

Страница 5: ...indicator denotes the battery bank is fully charged while a red indicator designates the battery is currently charging When the device is in operation charger is not connected the A and B LEDs are not...

Страница 6: ...on all headstages should be connected together to form a single common ground See Headstage Connection Guide on page 6 99 Distortion typical 1 A D Sample Rate Up to 48828 125 Hz Sample Delay Dependen...

Страница 7: ...me Description Pin Name Description 1 A1 Analog Input Channels 14 V Positive Voltage 1 5V 2 A2 15 GND Ground 3 A3 16 GND 4 A4 17 V Negative Voltage 1 5V 5 Ref Reference 18 HSD Headstage Detect 6 HSD H...

Страница 8: ...6 10 System 3 PZ2 PreAmp...

Отзывы: