Setup

 

Option

 

Notes

 

Active Channels 

Single or Four Channel 

In single channel mode, unused channels are shorted to ground 

Sample Freq 

3 kHz (1500), 6 kHz (3 kHz), 
12 kHz (6 kHz), System Rate 

System Rate is the processor’s rate up to 25kHz 
The number in ( ) represents the frequency content that can be 
acquired at the Sample Freq without aliasing. 

Impedance Frq 

100 Hz, 300 Hz, 500 Hz, 1 kHz 

This is the probe frequency for checking the impedance of 
electrodes 

Highpass Filter 

Off, 1 Hz, 3 Hz 

‘Off’ is a 0.1 Hz filter. There is a fixed 0.3Hz high pass analog 
filter on the inputs 

Auto Shutdown 

Never, 1 hour, 3 hours, 5 hours 

After a period of inactivity, the Medusa4Z will shut itself off to 
preserve power. Press Power to re-energize after Auto Shutdown.  

 

Maximum Voltage In 

+/- 10 mV 

S/N (typical) 

80 dB 

Input Referred Noise 
(μVrms) 

Depends on Medusa4Z sample rate and electrode impedance 

Medusa4Z Rate 

Shorted 

3 kOhm 

10 kOhm 

3 kHz 

0.15 

0.3 

0.5 

6 kHz 

0.25 

0.4 

0.7 

12 kHz 

0.3 

0.6 

1.0 

25 kHz 

0.4 

0.9 

1.4 

 

Sample Delay 

Dependent on Medusa4Z and RZ processor sample rates 

 

(RZ at 25 kHz)         

(RZ at 12 kHz)                           

(RZ at 6 kHz) 

Medusa4Z rate 

Samples 

Samples 

Samples 

25 kHz 

24 

N/A 

N/A 

12 kHz 

42 

24 

N/A 

6 kHz 

76 

42 

24 

3 kHz 

146 

76 

42 

 

Battery 

Li-Poly Battery 5 Ah capacity. Single channel mode, 30 hours between charges. Four 
channel mode, 24 hours between charges. 1000 cycles of charging, not removable by user.  

Medusa4Z_Input

Output-1

Output-2
Output-3
Output-4

In RPvdsEx (OpenEx, Synapse) 
 
Use the Medusa4Z_Input Macro. Set the number of desired channels. 
The macro does not account for sample delays.  

 

In BioSigRZ:  
Setup 

 Acquisition 

 Setup Inputs 

 Source 

 

Medusa4Z 
To avoid sample delays in your data trace, you must 
match acquisition rate in BioSigRZ to the Medusa4Z 
hardware Sample Freq (cycle Setup 

 Sample Freq 

 cycle Option). 25 kHz in BioSigRZ will be 

System Rate on Medusa4Z. 
* 12kHz recommended for ABRs  
Note: This software setting does not set the hardware 
sampling rate.  

  Important

: You must use TDT Software (Drivers/RPvdsEx or BioSigRZ) 

version 92 or later to use the Medusa4Z 

 

Software Configuration 