8. Configuration
The device can be configured via jumpers or using the additionally available USB
interface with the corresponding software “Digalox
®
Manager”. When configuring via
jumpers the basic settings can be made. The configuration via USB allows full
functionality.
Configuration via jumpers
Select measurement mode with jumpers at J4-J6 according to table "measurement
mode”.
Table measurement mode (factory preset, can be changed by software)
Measurement mode
Display
Selection
1
V AC direct
Volt AC
J1
J6
2
A AC direct
Ampere AC
J1
J6
3
V DC direct
Volt DC
J1
J6
4
A DC direct
Ampere DC
J1
J6
5
AC Frequency
Freq. Hz
J1
J6
6
5A AC scaled (current transformer)
Ampere CT
J1
J6
7
1A AC scaled (current transformer)
Ampere CT
J1
J6
To configure the upper scale caption, connect J3
within 10 seconds after restart and disconnect it
again. The device switches to programming mode.
“Prog” appears in the display, followed by the upper
scale caption, whereby the digit to be changed is
shown as “x”. Use J2 to select which digit is to be
changed. Change the value of the digit with J1.
Connect J3 again to save.
If the measurement mode "5A AC scaled" or "1A AC scaled" is selected the upper
scale caption must correspond to the primary value of the current transformer for
correct scaling of the measured value.
Configuration via USB interface with software “Digalox
®
Manager”
Connect the device and the computer using a USB cable. The device driver is
installed automatically if the computer is connected to the Internet.
After restart, the device always shows the first display value or
the one that is selected by the jumpers J4-J6 (see table
“measurement mode”).
-----
5 / 8
-----