5.5  r

eplacing

 

The

 

indenTer

When the hardness tester cannot be calibrated correctly anymore, the indenter is worn and must be replaced. 

•  Undo the pivot screw and take out the bottom handle. 

•  Take out the dial head but leave the cylinder in the frame. You can now see the adjusting nut. 

•  Loosen the adjusting nut (see also figure 4) and take out the indenter. 

•  Insert the new indenter and fasten the adjusting nut. Do not fasten the nut too tight as this will damage the 

indenter. 

•  Calibrate the hardness tester as described in paragraphs 5.3 and 5.4.

6.  m

ainTenance

This hardness tester is a precision instrument. Its service life depends on correct use and good maintenance. 

•  Clean the hardness tester every time you have used it with a soft cloth. 

•  The various parts of the tester have been treated to protect them against rust. However, always check the tester 

for rust when you use it often or when you are using it in a damp environment. 

•  Never drop the hardness tester. This will influence the correct working of the tester and may even render it 

useless.

•  Do not remove any parts of the tester other than for calibration purposes. Removing further parts voids the 

guarantee. 

7.  f

acTors

 

ThaT

 

will

 

inflUence

 

The

 

accUracy

 

of

 

The

 

TesTer

7.1  m

aTerial

The surface of the material to be measured needs to be free of dust and grains to ensure the accuracy of the 

measurement. 

7.2  d

isTance

 

To

 

The

 

edge

 

of

 

The

 

maTerial

The distance between the tip of the indenter and the edge of the material has to be at least 5 mm. When you 

measure closer to the edge this may influence the result. 

7.3  d

isTance

 

beTween

 

Two

 

measUring

 

poinTs

The distance between two measuring points has to be at least 6 mm. When you measure too close to an earlier 

measuring point this may influence the result.

7.4  o

xidaTion

Even though a layer of oxidation is thin, it will influence the result of a measurement. A layer of 10 µm will result 

in a deviation of 0.5 to 1 HW (over).

7.5  c

oaTing

A coating on the material will give you a unreliable result. Therefore you should always remove any coatings with 

sanding paper or a solvent. 

7.6  (i

n

)

correcT

 

Use

•  You should check the accuracy of the hardness tester regularly and calibrate it if necessary (see chapter 5). 

•  Always place the material horizontally between the tip of the indenter and the anvil. Make sure the material 

presses well against the anvil. 

•  Press down the handle firmly in one go. When you add pressure slowly this will result in an inaccurate reading. 

•  Make sure the material does not shift while measuring. 

Содержание HT MOD B

Страница 1: ...TCE HT MOD B H A R D N E S S T E S T E R ...

Страница 2: ... 4 5 2 Verification with test block 4 5 3 Calibration of the dial 4 5 4 Calibration of the load spring 4 5 5 Replacing the indenter 5 6 Maintenance 5 7 Factors that will influence the accuracy of the tester 5 7 1 Material 5 7 2 Distance to the edge of the material 5 7 3 Distance between two measuring points 5 7 4 Oxidation 5 7 5 Coating 5 7 6 In correct use 5 8 Conversion table and charts 6 ...

Страница 3: ...key 10 Indenter 5 Adjusting screw 11 Anvil 6 Dial head 12 Return spring 1 1 Indenter Figure 2 2 Specifications Testing range 0 20 HW Accuracy 0 5 HW Weight 0 5 kg 3 Application The TCE HT Mod B is used for measuring the hardness of aluminium and aluminium alloys within the range of 25 110 HRE 58 131 HV The thickness of the material to be measured can be between 0 4 8 mm and with an inner diameter ...

Страница 4: ...he case you must calibrate the hardness tester as described in paragraph 5 4 5 3 Calibration of the dial Figure 3 Press down the indenter firmly with the handle whilst there is no material between the indenter and the anvil Turn the adjusting screw with the screw driver to the left or the right until the pointer indicates 20 HW When the dial cannot be calibrated any more you must replace the inden...

Страница 5: ...parts voids the guarantee 7 Factors that will influence the accuracy of the tester 7 1 Material The surface of the material to be measured needs to be free of dust and grains to ensure the accuracy of the measurement 7 2 Distance to the edge of the material The distance between the tip of the indenter and the edge of the material has to be at least 5 mm When you measure closer to the edge this may...

Страница 6: ...6 CHANGES AND TYPING ERRORS RESERVED 8 Conversion table and charts Table 1 Figure 5 Figure 6 ...

Страница 7: ...sibility for defects which result from not reading the manual carefully or wrong use of the machine No rights can be derived from this manual All rights reserved No part of this manual may be reproduced in any form by print photoprint microfilm or any other means without written permission from the publisher TCE Tools BV Kennedylaan 14 Veghel the Netherlands Internet www tce tools nl ...

Страница 8: ...2016 V1 ...

Отзывы: