For a correct use of the appliance, we suggest that you do not fill the tray
with too many instruments to grant a good and uniform circulation of air.
Transport and storage
This hot-air steriliser is delivered with its original package and is considered
as FRAGILE good, as indicated on the cardboard box. In case of restitution
for repair, please use its original package. In case of storage, keep the
device away from humidity to safeguard its electrical components.
Maintenance and cleaning
Before each cleaning operation, unplug the device and let the machine cool
down first.
Clean the inside of the sterilisation chamber DAILY with a cloth moistened
with denaturated alcohol (90%).
Not do spray water, do not let water penetrate the chamber to avoid
electric shocks.
Check once a month that the supply cable is in good conditions.
In case of bad functioning, contact the Manufacturer.
In case of repair,
we recommend that you send back the device to the
Manufacturer with its original package.
Check periodically / yearly the temperature with the endowed pre-
set thermometer.
If problems arise, verify periodically that the safety device (thermostat) is
functioning.
If the safety device (thermostat) goes on,
unplug the oven
and press the
thermostat’s red button.
Electrical scheme
TH THERMOSTAT
L1 Funtioning bulb
L2 thermostat bulb
T timer
R resistance 230V 620W
F Fuses (2) 6.3x32 3.15A
TL line chuck
E earth chuck
TS safety thermostat