For  a  correct  use  of  the  appliance,  we  suggest  that  you  do  not  fill  the  tray 
with too many instruments to grant a good and uniform circulation of air. 

 

Transport and storage 

 

This hot-air steriliser is delivered with its original package and is considered 
as  FRAGILE  good,  as  indicated  on  the  cardboard  box.  In  case  of  restitution 
for  repair,  please  use  its  original  package.  In  case  of  storage,  keep  the 
device away from humidity to safeguard its electrical components.  

 

Maintenance and cleaning 

 

Before each cleaning operation, unplug the device and let the machine cool 
down first.  
Clean  the  inside  of  the  sterilisation  chamber  DAILY  with  a  cloth  moistened 
with denaturated alcohol (90%).  

Not do spray water, do not let water penetrate the chamber to avoid 
electric shocks.  

Check once a month that the supply cable is in good conditions.  
In case of bad functioning, contact the Manufacturer.  
In  case  of  repair,

 

we  recommend  that    you  send  back  the  device  to  the 

Manufacturer with its original package.  

 
Check  periodically  /  yearly  the  temperature  with  the  endowed  pre-
set thermometer.

  

If  problems  arise,  verify  periodically  that  the  safety  device  (thermostat)  is 
functioning. 
If the safety device (thermostat) goes on, 

unplug  the oven

 and press  the 

thermostat’s red button. 

 
 
Electrical scheme 
 

TH THERMOSTAT 
L1 Funtioning bulb 
L2 thermostat bulb 
T timer 
R resistance 230V 620W 
F Fuses (2) 6.3x32 3.15A 
TL line chuck 
E earth chuck 
TS safety thermostat

 

 
 
 

Содержание MINI AUTOMATIC

Страница 1: ...TAU STERIL s n c Via Gorizia 6 a Fino Mornasco CO TAU STERIL MINI AUTOMATIC INSTRUCTIONS FOR USE CE 0426 ...

Страница 2: ...een manufactured according to the EC norms The usage of this hot air oven is limited to those functions for which it has been projected and realised The Manufacturer declines any responsibility on risks or dangers due to improper usage by operators Guarantee lasts 12 months starting form the invoice date INSTALLATION Verify that transport and or stocking have respected acceptable humidity conditio...

Страница 3: ... values documented and declared by the Manufacturer of the material and or instrument to be sterilised Suggested temperatures by hot air sterilizers range from 150 C to 180 C Important the operator must respect the suggested sterilisation temperatures The maximum adjustable temperature is 200 C The suggested sterilisation times range from a minimum of 30 minutes to a maximum of 45 60 minutes The m...

Страница 4: ...tion chamber DAILY with a cloth moistened with denaturated alcohol 90 Not do spray water do not let water penetrate the chamber to avoid electric shocks Check once a month that the supply cable is in good conditions In case of bad functioning contact the Manufacturer In case of repair we recommend that you send back the device to the Manufacturer with its original package Check periodically yearly...

Страница 5: ...ed only for the purpose it has been meant for sterilisation of surgical instruments by surgeons or paramedical staff It is necessary to take any possible precautions to avoid improper use by unauthorised personnel It is forbidden to modify or alter in any way the thermic and electric safety protections of the device or put them out of use It is also forbidden to try to repair the machine on your o...

Страница 6: ...6 BEWARE DANGER OF ELECTRICAL SHOCK BEWARE WEAR PROTECTIVE GLOVES AEE WASTE ACCORDING TO LOCAL RULES ...

Отзывы: