PROBLEM
WHAT TO DO
The battery light stays on and
the hand stops working.
The batteries have gone flat and need to be recharged.
The battery charge does not
last a full day.
Possible reasons:
• the batteries are not fully charged. A full charge takes
four hours
• the batteries have been in use for more than one year
• the ‘anti-slip’ feature is on - this uses a lot of battery.
Turn off this feature in the app if not required.
• the ‘close’ signal is being used unnecessarily and is using
the battery. Once the hand has closed and has a solid grip
on an object, cease the ‘close’ signal
• Bluetooth has been left on and is using the battery.
Turn off Bluetooth when you do not need it.
Contact your clinician if the problem continues.
If you experience any other issues, please contact your clinician for further advice.