Coapt ControlRoom

Use the ControlRoom software to adjust the following:

 

„

Validate correct wiring connections and cable polarity

 

„

Manual Test to confirm proper connections and operation

 

„

Confirm electrode-to-skin contact

 

„

Select and calibrate grip patterns and other prosthesis’ motions

 

„

Firmware and Software updates

TASKA HandCal or MyTASKA

HandCal or MyTASKA is only required to adjust the following settings:

 

„

Set the hand into economy mode or disable low battery alarm.

 

„

EMG button setting – Grip Triggers or All Sensors.

 

„

Run a diagnostic test.

 

„

Modifying custom grips.

 

„

Assigning grips to the hand button.

1. Welcome

4. Configure

3. Bluetooth

Normal operation

When the TASKA Hand Bluetooth 
is on, grip changes and other 
communication from the Coapt system 
are ignored.

Always ensure the Bluetooth light is off 
when powering the hand on and using 
the Coapt system. 

If Bluetooth is on during startup,  
press and hold the hand button until 
the blue light goes out, then power 
cycle the prosthesis.

2. Review

Example of connection guide for Coapt ‘H2’ system 
using TASKA 6-band Coaxial Plug.

Close

TASKA 
Power 
Switch

Digital

Open

1

A

2

2

Clinician guides

All Prosthesis Connection guides for 
TASKA + Coapt systems are available at:

coaptengineering.com/clinicians

This includes guides for:

 

„

TASKA + Coapt

 

„

TASKA + Wrist R Coapt

 

„

TASKA + Wrist R Elbow 
s Coapt

Refer to the specific product guides for 
operation of elbow and wrist systems 
with the Coapt system.

 

 

Take care to connect the Coapt digital line (A) and the TASKA Power Switch 
cable (1) into the correct plugs to avoid damaging the hand. 

5. Connect

Bluetooth on: this prevents grip changes, but the 
hand may still open and close as normal.

Before you get started

The TASKA Hand works seamlessly with the Coapt Gen1 or Gen2 systems.
Please read the following to ensure a successful startup for the TASKA Hand:

 

„

The Bluetooth on the TASKA Hand must be turned off at all times for correct 
operation of the Coapt system.

 

„

Electrode settings are automatically configured by the Coapt system.  
There is no need to modify electrode settings using the TASKA software.

 

„

The OK (custom1) grip is assigned to Pincer Grip in the Coapt system and must 
not be modified.

 

„

Do not configure Grip Triggers and Idle Return grips using the TASKA software.  
 These will cause unpredictable operation with the Coapt system.

 

„

A 6-band Coaxial Plug is required when using the Coapt system with a Quick 
Disconnect TASKA Hand.

 

„

All Low Profile TASKA Hands come with a digital connection plug (labelled 
COMMS only). No additional adapters are required.

Bluetooth off: the Coapt system functions 
normally with the TASKA Hand, including driving 
grip changes.

Connecting to HandCal or MyTASKA

1.  Ensure HandCal is installed and the Bluetooth dongle is connected to the PC.*

2.  Hold down the hand button to turn Bluetooth on. 

3.  Start the HandCal software or MyTASKA app and connect to the TASKA Hand.

4.  Adjust the required settings as per section 

4. Configure**.

5.  To disconnect from the hand, click disconnect in the software.

6.  Turn the hand Bluetooth off by holding down the hand button.

7.  Power cycle the hand to reset Coapt.

Once reset, the Coapt system will operate as normal.

As the Coapt system automatically sets most essential settings in TASKA Hand, 
connecting to HandCal or MyTASKA for the initial setup may not be necessary.

* Not required when using MyTASKA.

  **  If creating or modifying custom grips, do not modify the OK grip (custom1), 

this grip is assigned to the Pincer grip within the Coapt system.

Содержание Coapt Gen1, Coapt Gen2

Страница 1: ...TASKA HandGen1 Gen2 with Coapt Gen1 or Gen2 System V1 1 Guide for Clinicians ...

Страница 2: ...he Coapt system There is no need to modify electrode settings using the TASKA software The OK custom1 grip is assigned to Pincer Grip in the Coapt system and must not be modified Do not configure Grip Triggers and Idle Return grips using the TASKA software These will cause unpredictable operation with the Coapt system A 6 band Coaxial Plug is required when using the Coapt system with a Quick Disco...

Страница 3: ...Coapt systems are available at coaptengineering com clinicians This includes guides for TASKA Coapt TASKA Wrist Rotator Coapt TASKA Wrist Rotator Elbow systems Coapt Refer to the specific product guides for operation of elbow and wrist systems with the Coapt system Take care to connect the Coapt digital line A and the TASKA Power Switch cable 1 into the correct plugs to avoid damaging the hand ...

Страница 4: ... off when powering the hand on and using the Coapt system If Bluetooth is on during startup press and hold the hand button until the blue light goes out then power cycle the prosthesis Bluetooth on this prevents grip changes but the hand may still open and close as normal Bluetooth off the Coapt system functions normally with the TASKA Hand including driving grip changes ...

Страница 5: ...o skin contact Select and calibrate grip patterns and other prosthesis motions Firmware and Software updates TASKA HandCal or MyTASKA HandCal or MyTASKA is only required to adjust the following settings Set the hand into economy mode or disable low battery alarm EMG button setting Grip Triggers or All Sensors Run a diagnostic test Modifying custom grips Assigning grips to the hand button 4 Configu...

Страница 6: ...from the hand click disconnect in the software 6 Turn the hand Bluetooth off by holding down the hand button 7 Power cycle the hand to reset Coapt Once reset the Coapt system will operate as normal As the Coapt system automatically sets most essential settings in TASKA Hand connecting to HandCal or MyTASKA for the initial setup may not be necessary Not required when using MyTASKA If creating or mo...

Страница 7: ... When this is illuminated the hand will only open and close This must be off for normal operation Hand button Can only be used if the EMG button is pausing Grip Triggers Hold to turn Bluetooth on off EMG indicator light Light flashes when Coapt signals are being limited Grip Triggers or All Sensors Fault light Light flashes if there is a fault Restart the hand Battery light Light flashes when the ...

Страница 8: ...ds The hand will still open and close freely 1 Press the EMG button This will pause grip changes from Coapt 2 On your TASKA Hand select the desired grip using the Hand Home buttons The hand will open and close as normal 3 Once the object is clear of the hand press the EMG button again to turn Coapt grip changes back on 2 3 1 ...

Страница 9: ...gnore all inputs from the Coapt system The light above the Hand button will flash and a beep may be heard 2 Input from the Coapt system is indicated by a green flashing light The hand will stay locked in position 3 Press the EMG button to unlock the hand and resume inputs from the Coapt system The light above the EMG button will flash and a beep may be heard 2 3 1 ...

Страница 10: ...g used if the hand is a quick disconnect model 4 Check the correct Coapt System is being used 5 Check the power and digital cables are correctly connected The hand changes grips randomly This might mean An idle grip and or Grip Triggers have been set in HandCal or MyTASKA Connect to the hand and remove these grips The calibration is not complete Re calibrate the Coapt system using the calibration ...

Отзывы: