INSTALLATION AND FUNCTIONING OF THE DEVICE
1) Do not install and do not try to repair the product if it was damaged during transport.
2) Do not connect the power cord to the power supply, unless specifi cally required.
3) In order to ensure the correct working of the device in safe conditions, it is absolutely forbidden to expose the
product to atmospheric agents and to direct heat sources. The temperature of the installation environment
has to be between 5 °C and 38 °C (41 °F – 100 °F).
4) Choose the place foreseen for the installation of the device. It must have the following features:
- It must have openings that allow change of air with the outside.
- The air change must be such to dissipate the heat of the device.
- Make sure that the room temperature does not exceed 38 °C (100 °F) during operation.
5) Place the device by taking into consideration the minimum recommended distances to allow visibility:
6) Lay the tubes as follows:
Insert the fi tting (supplied in the package, including gasket and
ring) inside the pipe, screw clockwise the fi tting orienting it as
needed and tighten fi rmly.
7) In connecting the pipes, check that the pipe coming from the
pump and/or from the fi lter group (not supplied as standard)
is connected in the position indicated with INLET (Ref. C) and
that the return pipe to the aquarium is connected in the posi-
tion indicated with OUTLET (Ref. D)
8) Start the pump and make sure that the water circulates regu-
larly within the circuit without any leaks. In case of abnormali-
ties of the hydraulic circuit or leaks thereof, inspect the con-
nections.
9)
WARNING: Make sure that the water that reaches
the device is fi ltered.
10) Make sure that the electrical power supply requirements cor-
respond to those shown on the technical data plate, applied on
the rear of the device.
11) With the pump in operation, insert the electrical power sup-
ply cable in the electrical current socket and by displacing the
electrical switch in position “1”, your device will start operating.
The temperature of water is indicated on the display. To view
the temperature to be reached, press the SET key; to return to
the water temperature, press once again the SET key or wait
for 5 seconds.
12)
WARNING: A 2-minutes delay was added to prevent
compressor failure at the fi rst start.
13) To stop the device, displace the electrical switch in position “0”.
WARNING: Only TK9000 and TK9000H. Connect to the electrical mains using a circuit breaker as
required by current applicable standards. The equipment shall be connected to a supply having
a system impedance lower than 0,24 Ω.
EN
Air outlet side (Ref. B)
>500 mm – 19.7 in
Filter side (Ref. A)
>300 mm – 11.8 in
Electrical power supply cable
side (Ref. F)
>60 mm – 2.3 in
Tab. 5-1