3
The automatic shut-down function shuts off the power automatically in the following cases:
• If you interrupt the measurement process. The power will shut down automatically within 30 seconds, depending upon the type of
operation.
• During programming, if you do not touch any of the keys or buttons within 60 seconds.
• After you have completed the measuring process.
To ensure accuracy, readings should be taken without clothing and under consistent conditions
of hydration. If you do not undress, always remove your socks or stockings, and be sure the
soles of your feet are clean before stepping on the measuring platform.
Be sure that your heels are correctly aligned with the electrodes on the measuring platform.
Don’t worry if your feet appear too large for the unit - accurate readings can still be obtained if
your toes overhang the platform.
It is best to take readings at the same time of day. Try to wait about three hours after rising,
eating, or hard exercise before taking measurements.
While readings taken under other conditions may not have the same absolute values, they
are accurate for determining the percentage of change as long as the readings are taken in a
consistent manner. To monitor progress, compare weight and body fat percentage taken under
the same conditions over a period of time.
Note:
An accurate reading will not be possible if the soles of your feet are not clean, or if your
knees are bent or you are in a sitting position.
You can change the weight unit by using the button on the back of the scale.
Press the Weight-only button when the display is blank. When “0.0” appears in the screen, press the Weight Mode button on the
underside of the unit to switch the unit in which weight is displayed.
Note:
If weight mode is set to pounds or stone-pounds, the height programming mode will be automatically set to feet and inches.
Similarly, if kilograms is selected, height will be automatically set to “cm”.
AUTOMATIC SHUT-DOWN FUNCTION
GETTING ACCURATE READINGS
CHANGING THE WEIGHT MODE
Note:
For use only with models containing the weight mode button
Heels centered on electrodes