background image

- 2 - 

 

 

Select  the  correct  wood  type  with  the  WOOD  TYPE 
knob  and  set  the  TEMPERATURE  COMPENSATION 
knob  to  the  appropriate  (measured  or  approximated) 
wood temperature. 

 

2.3. 

Conducting wood moisture cont. measurements 

 
To conduct measurements: 

 

Beat the electrode into the wood. A line drawn between 
the  needles  should  be  perpendicular  to  the  fibers  (the 
resistance should be measured across the fibers). 

 

Press the power button located near the BNC socket. 

 

Read the result. 
(If the moisture contents is lower than 6% a sign „LO” 
appears  on  the  LCD,  if  the  moisture  contents  is  higher 
than 99.9% - „HI” is displayed.) 

 
 

3. 

 

Concrete moisture contents 

 

3.1. 

Technical data 

 
 

The  measurement  of  moisture  contents  in  concrete  is 

based  on  the  measurement  of  the  resistance  and  complies 
with the British Standard BS.5352;1983. 
 
 

Measuring range 

1% - 10% 

 

Accuracy 

 

    within the range 

 

    of  0% - 5% 

0.5% 

 

    above 5% 

1% - 1.5% 

 
 

The  resistance  characteristics  have  been  selected 

during research and development. 

- 3 - 

 

3.2. 

Preparing the instrument for measurements 

 
 

Before  the  measurement  drill  two  holes  in  the 

concrete  floor 

  10  mm  in  diameter  and  40  mm  deep.  The 

distance  between  the  holes  should  be  10  cm.  Put  the 
electrodes  into the holes and fasten them by tightening their 
nuts. The torque applied to the nut should be appropriate to 
the  hardness  of  the  concrete  floor.  The  older  the  concrete 
floor (harder) the higher the torque can be. Do not crumble 
the  concrete  floor.  The  measurements  should  be  conducted 
in several places of the floor. 
 

3.3 

Conducting concrete moisture cont. measurements 

 
To conduct measurements: 

 

Connect the electrodes with the instrument using a cable 
with appropriate connectors. 

 

Set the WOOD TYPE knob to “CONCRETE”. 

 

Set The TEMPERATURE knob to 20°C. 

 

Press the power button located near the BNC socket. 

 

Read the result. 
(If the moisture contents is higher than 10% a sign „HI” 
appears on the LCD) 

 
Warning: 

 

For  concrete  moisture  contents  measurements  the 
TEMPERATURE COMPENSATION knob 

must

 be set 

to 

20°C

 

With  high  moisture  contents  values  (above  6  -  7%)  the 
result  may  “float”  a  little  –  it  will  slowly  decrease.  The 
result  observed  1.5  minutes  after  turning  the  device  on 
should be considered as the final result. 

- 4 - 

 

4. 

 

Air humidity 

 

4.1. 

Technical data 

 
 

The measurement of air humidity is conducted  using 

a chemical sensor (cellulose compound). This allows for an 
approximate measurement of air humidity in storage rooms, 
production houses etc. 
 
Measuring range 

30% - 90% RH 

Accuracy 

3% - 4% 

 

4.2. 

Preparing the instrument for measurements 

 
 

Place  the  SC1  sensor  in  the  room  where  the  air 

humidity  should  be  measured  at  least  8  hours  before  the 
measurement. 
 

4.3. 

Conducting the air humidity measurement 

 
To conduct the measurement: 

 

Plug the SC1 sensor into the instrument. 

 

Set the WOOD TYPE knob to “AIR HUMIDITY”. 

 

Set The TEMPERATURE knob to 20°C. 

 

Press the power button. 

 

Read the result when it stabilizes. 
(If the moisture contents is lower than 30% a sign „LO” 
appears  on  the  LCD,  if  the  moisture  contents  is  higher 
than 99.9% - „HI” is displayed.) 

 
Warning: 

 

For  air  humidity  measurements  the  TEMPERATURE 
COMPENSATION knob 

must

 be set to 

20°C

 