21
5.
OVERALL PERFORMANCE SPECIFICATION
5.1
Insulation resistance at 1000V DC
Not less than 10 megohms between mains input and frame.
5.2
Applied voltage test
2kV RMS for 1 minute between mains input and frame
5.3
Accuracy of instruments
KV meter ±1.5% at full scale
mA meter ±1.5% at full scale
5.4
H V Transformer
2.5kV RMS for 1 minute between primary and earth
2.5kV RMS for 1 minute between earthy end of secondary and earth
5.5
Complete Equipment
Over voltage test
Equipment run for 5 minutes at 115% of normal output voltage
Flash over-test
6 flash-overs direct to ground at 100% of normal output voltage