INDICATIONS FOR USE

The VariSoft

 infusion set is indicated for the subcutaneous infusion of insulin administered by an external pump. The VariSoft 

cannula and tubing make up the infusion set.

CONTRAINDICATIONS

VariSoft is neither intended nor indicated for use with blood or blood products (I.V. infusion).

WARNINGS

•  VariSoft infusion sets with t:lock tubing connectors should only be used with Tandem cartridges featuring the t:lock 

connector.

•  The VariSoft infusion set is only sterile and non pyrogenic if the packaging is unopened and undamaged. Do not use it if the 

packaging is already opened or is damaged.

•  Go through instructions for use carefully before inserting VariSoft, as failure to follow instructions may result in pain or injury.
•  Inaccurate medication delivery, infection and/or site irritation may result from improper insertion and maintenance of  

infusion site.

•  When using VariSoft for the first time, do so in the presence of a healthcare provider.
•  VariSoft is a single-use item, which must be disposed of after usage. Do not clean or re-sterilize.
•  Be sure that the needle guard is removed before insertion.
•  People with small or large amounts of subcutaneous fat should choose the insertion angle carefully, as the cannula could 

be placed in underlying muscle or dermal layer and hereby cause reduced or blocked medication delivery. Normal insertion 

angle between 20-45°. Consult your healthcare provider concerning this matter.

•  Do not leave air in the infusion set. Make sure to fill VariSoft completely.
•  Do not in any way put disinfectants, perfumes, deodorants or other products containing alcohol or disinfectants in contact 

with the tubing connector and the tubing, as these may affect the integrity of the infusion set.

•  Replace the infusion set if the adhesive tape becomes loose.
•  Check the infusion set frequently to ensure that the soft cannula remains firmly in place, and replace if it is not secured. Since 

the cannula is soft, it will not cause any pain if the soft cannula becomes fully or partially dislodged from the skin, and this 

may take place without notice. The soft cannula must always be completely inserted to receive full amount of medication.

•  Replace the infusion set every 48-72 hours, or per your healthcare provider’s instructions.
•  If infusion site becomes inflamed, replace set and use new site until the first site has healed.
•  Do not re-insert introducer needle into infusion set. Re-insertion could cause tearing of soft cannula which would result in 

unpredictable medication flow.

•  Never prime the infusion set or attempt to free a clogged tubing while set is inserted. You may accidentally inject too  

much medication.

•  Use aseptic techniques when temporarily disconnecting VariSoft, and seal the cannula housing with the cover provided. 

Consult with healthcare provider on how to compensate for missed medication when disconnected.

•  Protect VariSoft from direct sunlight. Store at room temperature.
•  Reuse of the infusion set may cause infection, site irritation, or damage to the cannula/needle. A damaged cannula/needle 

may lead to inaccurate medication delivery.

•  When infusing insulin, carefully monitor blood glucose levels when disconnected and after reconnecting.
•  When infusing insulin, and blood glucose level becomes unexplainably high, or occlusion alarm occurs, check for clogs and 

leaks. If in doubt, change infusion set since soft cannula could be dislodged, crimped or partially clogged. Discuss plan for 

rapid replacement of insulin with healthcare provider should any of these problems arise. Test blood glucose level to make 

sure problem is corrected.

•  When infusing insulin, check the blood glucose level 1-2 hours after introducing VariSoft and check the infusion site several 

times a day, and measure your blood glucose regularly.

•  When infusing insulin, do not change infusion set just prior to bedtime, unless blood glucose can be checked 1-2 hours  

after insertion.

RECONNECTING

DISCONNECTING

Consult pump 

manufacturer’s 

instructions for use 

when connecting 

infusion set to  

insulin pump.

CONTENTS

A. 

Needle guard

B. 

Soft cannula

C. 

Adhesive tape

D. 

Introducer needle

E. 

Cannula housing

F. 

Tubing

G. 

Connector needle 

H. 

Circular protective cap

I. 

Disconnect cover

Use by (year-month-day)

Do not re-use

Sterilized using ethylene oxide

Attention: See accompanying documents

See Instructions for Use

Reference / model number

Lot number

Manufacturer

© 2017 Tandem Diabetes Care, Inc. All rights reserved. Tandem Diabetes Care is a registered 
trademark and VariSoft is a trademark of Tandem Diabetes Care, Inc. AW-1000226_A

Unomedical a/s

Aaholmvej 1-3, Osted  

4320 Lejre, Denmark
Made in Denmark

Legal manufacturer:

Distributed by:

11045 Roselle Street 

San Diego, CA 92121, USA

www.tandemdiabetes.com 
+1-858-255-6269

Fill the empty space in 

the cannula with the 

following prime:

U-100 insulin:

0.7 units (0.007 ml)

Instructions for Use