Chapter 20 – Technical Specifications
…table continued from previous page
Specification Type
Specification Details
Basal Delivery Accuracy at all Flow
Rates (tested per IEC 60601-2-24)
±5%
The Pump is designed to vent automatically when there is a pressure difference between inside the
cartridge and the surrounding air. In certain conditions, such as a gradual elevation change of 1,000
feet, the pump may not vent immediately and delivery accuracy can vary up to 15% until 3 units have
been delivered or elevation changes by more than 1,000 feet.
Bolus Delivery Accuracy at all Volumes
(tested per IEC 60601-2-24)
±5%
Patient Protection from Air Infusion
The pump provides subcutaneous delivery into interstitial tissue and does not deliver intravenous injections.
Clear tubing aids in detecting air.
Maximum Infusion Pressure Generated
and Occlusion Alarm Threshold
30 PSI
Frequency of Basal Delivery
5 minutes for all Basal Rates
Retention Time of Electronic Memory
when Internal Pump Battery is
Fully Discharged (including Alarm
Settings and Alarm History)
Greater than 30 days
Infusion Set used for Testing
Unomedical Comfort
™
Infusion Set
table continued on next page …
Содержание t:slim X2
Страница 2: ......
Страница 14: ...Section 1 Before You Begin...
Страница 15: ...Chapter 1 Introduction...
Страница 19: ...Chapter 2 Important Safety Information...
Страница 31: ...Chapter 3 Getting to Know Your t slim X2 Pump...
Страница 54: ...Section 2 Key Pump Features...
Страница 55: ...Chapter 4 Getting Started...
Страница 60: ...Customer Technical Support 877 801 6901 60 Chapter 4 Getting Started THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 61: ...Chapter 5 Infusion Site Care and Loading Cartridge...
Страница 75: ...Chapter 6 Personal Profiles...
Страница 84: ...Customer Technical Support 877 801 6901 84 Chapter 6 Personal Profiles THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 85: ...Chapter 7 Bolus...
Страница 92: ...Customer Technical Support 877 801 6901 92 Chapter 7 Bolus THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 93: ...Chapter 8 Stop Resume Insulin...
Страница 96: ...Section 3 Additional Pump Operations...
Страница 97: ...Chapter 9 Temporary Basal Rate...
Страница 100: ...Customer Technical Support 877 801 6901 100 Chapter 9 Temporary Basal Rate THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 101: ...Chapter 10 Quick Bolus...
Страница 104: ...Customer Technical Support 877 801 6901 104 Chapter 10 Quick Bolus THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 105: ...Chapter 11 t slim X2 Pump Settings...
Страница 108: ...Customer Technical Support 877 801 6901 108 Chapter 11 t slim X2 Pump Settings THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 109: ...Chapter 12 t slim X2 Pump Info and History...
Страница 111: ...Chapter 13 t slim X2 Pump Reminders...
Страница 116: ...Customer Technical Support 877 801 6901 116 Chapter 13 t slim X2 Pump Reminders THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 117: ...Chapter 14 User Settable Alerts and Alarms...
Страница 120: ...Section 4 t slim X2 Pump Safety Alerts and Alarms...
Страница 121: ...Chapter 15 t slim X2 Pump Alerts...
Страница 143: ...Chapter 16 t slim X2 Pump Alarms...
Страница 157: ...Chapter 17 t slim X2 Pump Malfunction...
Страница 160: ...Section 5 Living With and Caring for Your t slim X2 Pump...
Страница 161: ...Chapter 18 Lifestyle Issues and Travel...
Страница 164: ...Customer Technical Support 877 801 6901 164 Chapter 18 Lifestyle Issues and Travel THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 165: ...Chapter 19 Taking Care of Your t slim X2 Pump...
Страница 168: ...Section 6 Technical Specifications and Warranty...
Страница 169: ...Chapter 20 Technical Specifications...