Customer Technical Support:
(877) 801-6901
31
Personal Profile
Programming a New Profile
NOTE
Data entered will affect insulin delivery. Make sure all information is entered correctly.
1. Tap
Basal
.
NOTE
The first time segment will start at
12:00 AM (Midnight).
2. Using the on-screen keypad, enter
the desired basal rate and tap
DONE
.
NOTE
Always confirm that the decimal point
placement is correct.
3. Tap
Correction Factor
.
COLOR NOTE
The orange highlighted area indicates a
recent change made to a setting.
4. Using the on-screen keypad, enter
the desired correction factor and
tap
DONE
.
NOTE
Enter the mg/dL that 1 unit of insulin
will lower BG.
COLOR NOTE
Gray text indicates values that do not
need to be added.
12:00 AM
SAVE
BACK
Basal
Press to Set Up
Correction Factor
Press to Set Up
Carb Ratio
Press to Set Up
Target BG
Press to Set Up
u/hr
2
1
3
0
8
7
9
5
4
6
DONE
BACK
Enter Rate
.
12:00 AM
SAVE
BACK
Correction Factor
Press to Set Up
Carb Ratio
Press to Set Up
Target BG
Press to Set Up
Basal
0.75
u/hr
2
1
3
0
8
7
9
5
4
6
DONE
BACK
unit
mg/dL
--
1 :
Содержание t:slim X2
Страница 2: ......
Страница 6: ...Overview...
Страница 22: ...Getting Started...
Страница 30: ...Personal Profile...
Страница 47: ...Customer Technical Support 877 801 6901 45 Personal Profile THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 48: ...My CGM...
Страница 65: ...Customer Technical Support 877 801 6901 63 My CGM THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 66: ...Basal IQ Technology...
Страница 71: ...Customer Technical Support 877 801 6901 69 Basal IQ Technology THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 72: ...Load...
Страница 83: ...Customer Technical Support 877 801 6901 81 Load THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 84: ...Bolus...
Страница 102: ...Pump Features...
Страница 115: ...Customer Technical Support 877 801 6901 113 Pump Features THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 116: ...Alerts and Technical Support...
Страница 119: ......