![T&D TR-77Ui Скачать руководство пользователя страница 26](http://html1.mh-extra.com/html/tandd/tr-77ui/tr-77ui_user-manual_3096900026.webp)
23
Options
Temperature/Humidity Sensor-High-Precision Type
(HHA-3151)
5
unit: millimeters
Materials:
Temperature/Humidity Sensor
Polycarbonate
Vinyl chloride-shielded w
ire
Measurement Range: Temperature -30 to +80°C, Humidity 0 to 99%RH
Accuracy: Temperature ±0.3°C [at 0 to 50 °C], ±0.5°C [at all other temperatures]
Humidity ±2.5 %RH [at 25 °C, 10 to 85 %RH]
±4.0 %RH [at 25 °C, 0 to 10 % or 85 to 99 %RH]
At temperatures other than 25 °C and
0 °C, add ±0.1 %RH
per degree of difference from 25.
Humidity Hysteresis: ±1.5 %RH or lower (*1)
Responsiveness: Temperature Response Time (90%) Approx. 7 min.
Humidity Response Time (90%) Approx. 20 sec.
Long Term Stability: 1%RH/yr, ±0.1°C/yr
Conditions for Use: Do not expose to condensation, dampness, corrosive gases or organic
solvents.
Cable Length: 1.5 m
Sensor Extension Cable (TR-1C30)
unit: millimeters
Cable Length: 3.0 m
Materials:
Vinyl chloride-shielded wire
NOTE:
It is possible to use up to three extension cables per sensor.