Cautions about using the
I
lluminance-UV Sensors
Do not connect the ISA-3151 sensor to any data logger other than
the TR-74Ui.
Use the sensor within the operating environment specifications.
When measuring ultraviolet light or other light which may cause
damage or injury to your eyes or skin make sure to use protection
such as safety glasses or some type of light-proof shield.
Do not expose the sensor to a strong impact.
Cracks or scratches in the
I
lluminance sensor and/or in the UV sensor will
adversely affect the measurement accuracy. Also, a broken sensor may result in
injury.
This sensor is not waterproof. By all means do not allow it to get wet.
Do not use in any environment that is exposed to corrosive gas and
organic solvents. Also, do not use in areas near fire or exposed to
excessive heat.
When the
I
lluminance-UV sensor is not being used, please store it
in a place with a normal temperature and without condensation.
If the sensor surface gets dirty, wipe it with a soft cloth.
If the sensor surface accumulates impurities (dirt), it will cause a decrease in
the sensor’s accuracy and sensitivity.
If you wish to extend the length of the sensor cable, please purchase
our optional sensor extension cable separately.
Do not cut or process the sensor cables.
"Adjustment Settings" made in the supplied software are saved to
the sensor. Therefore, when a sensor is replaced, it is necessary to
re-make any desired adjustment settings before using the new
sensor.