Getting Ready and Downloading Data

1.

Using the software 

T&D Recorder for Windows

 which comes 

with the Data Collector TR-57DCi, make necessary settings for 
infrared communication function and set to 

Permit

. For more 

details, see the User

s Manual that comes with the software.

2.

With the Unit and the TR-57DCi face-to-face and with about 10 
to 20 cm of space between them, download the recorded data 
from the TR-51i/52i to the TR-57DCi(within 15 degrees from side 
to side and top to bottom).

For detailed information about the operation and functions of 

the Data Collector TR-57DCi, see the TR-57DCi User

s Manual. 

10

20cm 15

°

15

°

Set-up Procedure

1. 

As instructed by the software, connect the Communication 
Port or Data Collector to your computer using the provided 
communication cable.

Communication Cable

2.

Place the Unit face down on the Data Collector or 
Communication Port, making sure that the optical 
communication areas are aligned properly.

Data Collection Devices (Data Collector)

           TR-57U , TR-57DCi

           RTR-57U, RTR-57C

Data Collection Devices 
  (Communication Port)

     TR-50U

Recording settings for the Unit and the downloading of recorded 
data from the Unit can now be carried out from your computer.

 Notes about Optical Communication

-Proper communication may not be possible in the following situations: 

where temperatures are very high or very low, in an environment 
with intense brightness (higher than 5,000lx), or when the remaining 
battery life for the Unit is very low. 

Recording Settings for the Unit

Recording settings for the Unit can be changed by using the 
software supplied with the Communication Port or Data Collector.

The factory default settings are as follows: Recording Interval at 10 
minutes, Recording Start at Immediate Start, Recording Mode at Endless, 
Infrared Communication at Forbid.

Recording 
Intervals

Select from 15 choices: 
                   1, 2, 5, 10, 15, 20, and 30 seconds or 
                   1, 2, 5, 10, 15, 20, 30, and 60 minutes

Recording Start

Immediate Start     

Recording starts immediately upon battery   

                                 installation.
Programmed Start

Recording starts on the set date and time.

Recording 
Modes

One-Time

Upon reaching storage capacity of 16,000 
readings, recording automatically stops

Measurements and the word "FULL" will alternately 

appear in the LCD.

Endless

Upon reaching the storage capacity of 16,000 
readings, the oldest data is overwritten and 
recording continues.

Infrared 
Communication 
Function

Permit

Infrared Communication will be possible.

Forbid

Infrared Communication will not be possible.

It is possible to carry out the following data processing by using the software.

  

View and Print

View and print graphs and lists of recorded data. 

  Save                

Create and save recorded data files and text files.

 Notes about Communication Devices 

-When using a Data Collector, you can download recorded data, view the 
 data in graph form, and make all necessary recording settings without 
 connecting to a PC.
-The time necessary to download one Unit of full data varies depending 

upon the type of device being used.

Communication with your Computer

 In order to change settings in the Unit such as recording settings, download recorded data from the Unit to your computer, or communicate with your computer, it 

is necessary to purchase separately a Communication Port or Data Collector. 

 By using the Software "T&D Recorder for Windows

 supplied with the Communication Port or Data Collector, it is possible to carry out communication 

with a PC. For details about how to make recording settings, download data and other operations, please see the User

s Manual that comes with the 

Communication Port or Data Collector.

 The latest version of 

T&D Recorder for Windows

 can be downloaded free of charge from our Web Site. (TR-51i and TR-52i can be operated with the Software 

version 1.80E or later.)

-This Unit is an infrared-equipped device. Do not place the Unit in areas exposed to direct 

 sunlight, directly below an incandescent lamp, or near other infrared devices.

 Placing in such an area may cause infrared communication to not work properly. 

-Also, Infrared communication may not be carried out properly in a 

 low-temperature environment (below -20

).

-Proper communication may not be possible if the infrared port is dirty.

-During the infrared communication, do not touch the infrared port or 

 interrupt the communication.

-Before replacing a battery, please make sure to download any necessary data and proceed 

 with changing the battery.

-Downloading of data cannot occur while the battery is removed.

Cautions for Using in a Low-temperature Environment below -20

-When using in an environment below -20

the battery life will be halved.If the battery life 

 has depleted to half or less under normal conditions and then is moved to an environment 

 of –20

 or less, the battery warning mark will come on and in about one day

s time the 

 battery may lose all power. If you use the Unit in a low-temperature environment, we 

 suggest replacing the battery as early as possible.

Reading the LCD

Basic LCD Display

When being used in very hot or cold environments, the display may become difficult to 
read.

 

This is not a malfunction.

Recording ([REC] Mark) The recording status is shown here.

ON : Recording in progress or Storage Capacity FULL.
BLINKING :Waiting for programmed start.

Battery Life Warning Signal

When it is time for the battery to be replaced, this 
signal will appear.

Recording Mode

  :One-Time

 

  : Endless

 

Measurement and 
Message Area

Measurements or operational messages are 
shown here.

I

nfrared Communication 

([Ir] Mark)

The infrared communication status is shown here.
ON : Permitting infrared communication.
Not ON : Forbidding infrared communication

.

Battery Replacement Mark

1. 

When it is time for the battery to be replaced, a battery life 
warning signal will appear.

Please replace the battery as soon as possible if this mark appears.

2.

After removing the battery, wait for about three seconds 
until the 

bAtt

 mark appears. Once this appears, please  

insert the new battery as quickly as 

possible.

-If the battery is replaced before the 

bAtt

 mark appears, the

 battery life warning signal may remain even after replacing.
-If you change the battery at this point, all recorded data will be saved.

3.

If the battery is further left unchanged, the display will 
automatically shut off.

If, at this time, a new battery is placed in the Unit, the 

CHEC

 mark will 

appear on the display after which recording will begin again using the 
previously set recording conditions. Note however that all previously 
recorded data will have been lost.

Estimating Battery Life

The battery-life warning signal will appear based upon the calculation of 
battery use. This mark may not appear correctly if the same battery is taken 
out and put in, therefore do not remove the battery until it can be replaced 
with a new one. If infrared communication is set to be permitted, battery life 
will be shortened. 

When communication frequency is 4 times a month:

Recording Interval

1 second

2 seconds

5 seconds

10 seconds 
or longer

Infrared Communication  Forbidden

About 
18 months

About 
2 years

About 
3 years

About 
4 years

Infrared Communication  Permitted

About 
14 months

About 
14 months

About 
19 months

About 
2 years

- The battery life warning signal may appear sooner than noted above.
- Battery life will be shortened if used under the following conditions: downloading 

data very often, setting the recording interval at less than 10 seconds, leaving the 
Unit with the warning LED blinking, measuring in an environment below -20

, or 

leaving the Unit unconnected to the sensor for several months (TR-52i).

Other Messages

[FULL (Storage Capacity FULL)]

When Recording Mode has been set to 

One 

Time

 and the Unit reaches its storage capacity of 

16,000 readings, recording will automatically stop 
and in the LCD the measurement reading and the 
word 

FULL

 will alternately appear.

   Estimation of time until 

FULL

 is displayed

Recording 

Interval

1 second

30 seconds 1 minute

10 minutes 60 minutes

Period

About 4 

hours

About 5 

days

About 11 

days

About 111 

days

About 1 year 

and 10 months

[Check]

If this appears, all data that was stored in the Unit 
will have been erased.

This will be displayed under the following 
conditions:

- The first time a battery was inserted after purchase
- If the battery is replaced after having been taken out 

for a long period

- If the battery is replaced after the battery power has 

been lost.

Warning (set limit exceeded)

Warning LED (flashes)

Using the Software that comes with the 
Communication Port or Data Collector, you can 
make settings for the Upper / Lower Limits and 
Judgment Time. If a measurement exceeds one 
of the set limits, the warning LED and a message 
will be displayed.

[Upper Limit Exceeded]

If a measured temperature exceeds the set 
upper limit, the Logger LED will alternately flash 
between [Hi] and the measurement.

[

Lower Limit Exceeded]

If a measured temperature exceeds the set 
lower limit, the Logger LED will alternately flash 
between [Lo] and the measurement.

 

Starting the Warning Monitoring Function

If these settings are made in an environment where one of the limits is 
being exceeded and recording is started, the monitoring function will enter 

wait

 mode. Once a measurement returns to within the set limits, the 

monitoring function will begin to operate.

 

How to Turn Off a Warning

-Restart recording from the software.
-Use [Clear Warning] setting from the software (only via TR-50U).
-Download the recorded data (only when successfully completed).
-Produce a condition so that 

CHEC

 is displayed. (see [Check] above).

[Measurement Range Exceeded (for TR-52i only)]

The temperature display blinks when the 
temperature exceeds the measurement range 
(-60

 or below / 155

 or above).

[Sensor Unconnected (for TR-52i only)]

This will be displayed when a sensor has not 
been connected or the wire has been broken.
Measurement and recording will continue so 
battery power will be consumed.

 

-If after re-connecting the sensor and measurements can still not 

be displayed, it is very possible that the sensor or the Unit are 
defective or have been damaged.

Downloading Recorded Data via Infrared Communication

By using a Data Collector TR-57DCi, it is possible to download 
the recorded data from a TR-51i /52i via infrared communication.

Note

 

-

Please note that infrared communication is a function limited to the  downloading

      of recorded data. Other functions such as making Unit settings cannot be carried 
      out.

 

-

For downloading one TR-51i/52i Unit at full storage capacity, it takes about 60

      seconds.

 

Cautions about the Infrared Communication Sensor

    

Cautions about Changing the Battery