Regulatory Information | E2
Susceptibility and ESD Limits
The following table lists radiated, magnetic radiated, and
conducted susceptibility limits for the SDLT 600 tape drive.
Type
Specifications
Comments
Radiated immunity:
high frequency, electric fields
3 V/m (rms), 80%
modulated, 1 kHz,
26–1000 MHz
No errors
*
No screen distortion
* The SDLT 600 tape drive maintains normal operation both in Read/Write and in Standby
conditions. No errors attributable to the test are encountered.
Magnetic radiated:
low frequency, magnetic fields
3 A/m @ 230V/50 Hz
No errors*
No screen distortion
EFT: Fast transient (bursts)
for power and data cables
2 kV
No errors*
PLT: High energy transient
voltage for power cables
1.2 kV (Differential)
2.5 kV (Common Mode)
No errors*
Low-level conducted
†
interference for AC and
DC cables
† Conducted: The transient voltage is the actual peak voltage above the normal AC voltage
from the power source. The maximum energy in a single pulse from the transient generator
must be limited to 2.5 W.
3 V/m (rms),
80% modulated, 1 kHz,
0.15–80 MHz
No errors*
No screen distortion