damage if the respective packing material is not available
for examination).
The shipment contains at least the bath as mentioned in
the delivery checklist. Further the consignment might
contain one or more viscometers, individually packed in
small boxes with the calibration certificate included in the
box, as well as ASTM thermometers, thermometer
holders etc. Please see the packing list for details
concerning total contents of consignment.
Before filling the bath remove any remaining packing
material from its interior. The interior of the bath can
be accessed by taking off the lid on the top of the
bath.
4.3
Initial use
Before filling the bath make sure that the bath is clean.
Place the bath spirit level, and connect it to the mains
after the mains switch is placed in “off” position. Use a
mains supply with proper grounding.
Important: Never use the bath without fluid
4.4
Filling the bath
The bath-fluid level must be 10- to 20 mm below the edge
of the top-plate.
In no case the fluid level may be lower than 60 mm under
the edge. This can cause severe damage to the heating
element.
10 - 20 mm
Max. 60 mm