Tamson Instruments TC16 Скачать руководство пользователя страница 1

 

 

 

 

 

 

 

 

 

 

   

Van ’t Hoffstraat 12

 

2665 JL Bleiswijk, The Netherlands 
T. 31 (0) 10 522 43 73 

 

 
E-mail: [email protected] 
Website: www.tamson.com

 

 

Tamson Instruments bv 

Page 1/20 

 

tc16-40-58-man.docx Rev. 2.4 UK 9-21

 

ISO 9001 : 2015 
NL/PRO 238239125 

 

VAT: NL 80 66 34 984 B01 

Bank account no.:  

NL28 INGB 0007 350 370 

NL95 RABO  0160100046 

Chamber of commerce 27 16 95 41 

 
 
 
 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

USER-

MANUAL 

TC16-40-58

 

 

Содержание TC16

Страница 1: ...l sales tamson com Website www tamson com Tamson Instruments bv Page 1 20 tc16 40 58 man docx Rev 2 4 UK 9 21 ISO 9001 2015 NL PRO 238239125 VAT NL 80 66 34 984 B01 Bank account no NL28 INGB 0007 350...

Страница 2: ...TION TO THE TC SERIES 8 6 1 GENERAL 8 6 2 PUMP 8 6 3 COOLING 8 6 4 OVER TEMPERATURE SAFETY 9 6 5 ADJUSTING THE MECHANICAL OVER TEMPERATURE PROTECTION THERMOSTAT 9 6 6 FLUID LEVEL 10 6 7 FLUID LEVEL DE...

Страница 3: ...damage to the equipment or its accessories caused by incorrect operation Important When performing service maintenance or moving the apparatus always disconnect the apparatus at the main s socket Prop...

Страница 4: ...el Avoid damage to the equipment or its accessories through incorrect operation Environment Panel sealing Confirms EN60529 IP65 Environment Temperature 0 tot 35 C Supply enough ventilation Humidity 5...

Страница 5: ...ate included in the box as well as ASTM thermometers thermometer holders etc Please see the packing list for details concerning total contents of consignment Before filling the bath remove any remaini...

Страница 6: ...ting temperature When viscosity of the bath fluid is too high at room temperature the motor fuse will be activated In such a case the bath fluid can first be heated to set point temperature When the b...

Страница 7: ...370 NL95 RABO 0160100046 Chamber of commerce 27 16 95 41 Do not use demineralized or distilled water This water can lead to corrosion of the bath and moving parts The corrosion can wear out the pump b...

Страница 8: ...rough a closed external system The standard pump offers a motor protection small thermal fuse which is placed on the back panel of the motor compartment This protects the motor from excessive loads Wh...

Страница 9: ...l cause extreme temperature raise To prevent high temperatures the bath is fitted with a mechanical temperature switch thermostat This thermostat will switch off the bath at a preset temperature in th...

Страница 10: ...too low When the bath has been installed it must be filled with an appropriate liquid When working with water the bath should be filled to 1 cm below the lid For oil the bath should be filled to not...

Страница 11: ...g on If the bath has been properly filled with fluid it can be switched with the mains switch located on the front panel Choose a working temperature set point 7 2 Control Panel 1 LED Display 2 LED Ov...

Страница 12: ...7 Switch for fan only for TC16 with p n 31T0671 31T0861 8 Fan only for TC16 with p n 31T0671 31T0861 7 4 Selecting parameters The front panel layout shows the following four operating keys Press short...

Страница 13: ...D values can not be set manually anymore pb Changing the tuning PID parameters Proportional band value Default value is 1 range 1 100 Proportional band equals 1 proportional value Press page key to co...

Страница 14: ...ir own heat capacity the use of external cooling and external connected processes circulation or working at different set point temperatures requires new settings of the PID parameters These parameter...

Страница 15: ...elow provides fast finding of the PID settings and is refereed to as the method of Ziegler Nichols This procedure is only to be followed when automatic tuning is not functioning This method provides f...

Страница 16: ...nt temperature to remove the fluid make sure that the hose to drain and waste container can withstand the temperature of the bath fluid When removing bath fluid do not inhale toxic vapor Always use ap...

Страница 17: ...nstruments bv Problems with set point Heater LED is not burning motor is turning and temperature raises above set point Setpoint too near to room temperature Cooling of the bath is needed Temperature...

Страница 18: ...error Auto reset after 2 s E6 Watchdog error Auto reset after 2 s 10 SPECIFICATION 10 1 Technical specifications overview Item Unit TC16 TC40 TC58 Ordering code 230V 50 or 60Hz 00T0671 00T0681 00T069...

Страница 19: ...safe handling procedures should be provided to handlers and users Alteration of this document is strictly prohibited Except to the extent required by law republication or retransmission of this docum...

Страница 20: ...6 42 EC Machinery Directive of the European Parliament and of the Council of 17 May 2006 42 EC 2nd Edition June 2010 p d EN 60204 Machinery Directive and Safety requirements p pi e EN60950 1 Low Volta...