Talley ATTIVO Скачать руководство пользователя страница 6

5

OPERATION BUTTONS

The operation buttons on the face of the power unit provide the following functions.

RUN/STOP

Press to invoke stand-by mode prior to running power unit. Press again to run power unit. 

Press and hold whilst power unit is running (and unlocked) to cease operation and return 

to stand-by mode. In battery operation the power unit will then power off after 1 minute of 

inactivity. If using mains power, switch off by disconnecting power cable from power unit or 

turning off mains power.

COMFORT CONTROL

The automatic default comfort setting is MEDIUM. However, if the patient 

prefers a firmer or softer cushion, increase or decrease the comfort control 

setting accordingly using the UP and DOWN arrow buttons (SOFT/MEDIUM/FIRM). The comfort 

setting is shown on the display screen. Check periodically to ensure patient support and comfort.

MUTE

Press to silence the sounder and to clear the message from the display screen. The 

MUTE button can also be used to force power-off in stand-by mode.

DATA

The information menu can be activated by pressing the DATA button for 5 or more 

seconds while pressing the RUN/STOP button, then releasing the RUN/STOP button 

while continuing to press the DATA button for a few seconds. Press the DATA button when 

ready to exit the information menu.

NB. Used for accessing information only, does not affect mode of operation.

UNLOCK / LIGHT

The power unit will automatically lock 1 minute after last button operation when 

running to prevent inadvertent operation of button functions (except MUTE), as 

indicated by ‘   ’ on the display screen. Press and hold the UNLOCK button until power unit 

beeps if further button operation is needed (i.e. change of comfort setting, or returning 

power unit to stand-by mode). The power unit will lock again 1 minute after last button 

operation. Pressing this button will also illuminate the display screen for 10 seconds if 

power unit is in battery operation (screen will always be illuminated in mains operation).

MAXIMUM USER WEIGHT GUIDELINES

127kg (20 stone) max.

Содержание ATTIVO

Страница 1: ...ATTIVO Dynamic Seating System User Guidelines reducing avoidable harms ...

Страница 2: ...This is a statement that alerts the user to the possibility of serious injury or other adverse reactions with the use or misuse of the device This is a statement that alerts the user to the possibility of a problem with the system associated with its use or misuse IP Ingress Protection Pump only 2 Protection against fingers or other object not greater than 80mm in length and 12mm in diameter 2 Pro...

Страница 3: ...Power unit carry case l TS209 B A S E SEQUENTIAL cushion or B A S E Recliner Mat The following accessories spares are also available l 12V Vehicle adaptor charger l Spare mains lead with integral power supply 15V Mains adapter FW7362M 15 Cautions and Warnings CONTRAINDICATIONS FOR USE Alternating pressure therapy should not be used for patients with unstable fractures gross oedema burns or an into...

Страница 4: ...ct fuse used l The cushion power unit and power adapter should be cleaned between patient use please refer to Care and Maintenance section for all warnings and cautions l The power unit battery is installed inside the unit and is not accessible by users Battery replacement should only be carried out by qualified Talley service personnel l The power unit must only be used with the approved power ad...

Страница 5: ...e the battery is fully charged when needed A fully discharged battery will take a number of hours to fully charge 6 Standby mode is automatically invoked when the power is switched on Press the RUN STOP button to initialise and run the seating system When the power unit has finished initialising the pump screen will display ATTIVO together with the battery status and the comfort setting Fig 2 7 Th...

Страница 6: ...isplay screen The MUTE button can also be used to force power off in stand by mode DATA The information menu can be activated by pressing the DATA button for 5 or more seconds while pressing the RUN STOP button then releasing the RUN STOP button while continuing to press the DATA button for a few seconds Press the DATA button when ready to exit the information menu NB Used for accessing informatio...

Страница 7: ...acement should only be carried out by qualified service personnel Contact Talley or authorised dealer for battery replacement service Care and Maintenance COVER Always keep the cushion cover as clean as is practicable The material is waterproof and vapour permeable l Inspect top cover for signs of damage or wear which could result in the contamination of the interior e g tears holes damage to seam...

Страница 8: ...adapter in accordance with the local regulations including WEEE requirements The power unit power adapter should be cleaned between patient use as a minimum In line with the MHRA Medical Device Alert MDA 2013 019 Talley advises customers to use pH neutral high level disinfectant cleaning products to sanitise reusable medical devices to prevent damage to materials and the degradation of plastic sur...

Страница 9: ...ntee The intended design life is 5 years if fully serviced Fault Finding All sounders can be silenced and messages cleared by pressing the MUTE button once Should any fault occur press the MUTE button to reset the power unit If fault remains re occurs contact Talley WARNING INDICATORS NO BATTERY indicates battery is not correctly fitted The power unit will only operate using the power adapter whil...

Страница 10: ...mensions 430mm x 430mm x 70mm 1080mm x 430mm x 70mm Weight 1 9 kg 3 9 kg The above mains adapter is considered part of the ME equipment 12V vehicle power adapter is available as an optional accessory 99 01 14 05 The ATTIVO power unit must only be used with the specific external power adapters as supplied by Talley EMI fault indicates that the unit detects the pressure sensor amplifier is adversely...

Страница 11: ...nterfering equipment repaired or replaced l Reorient or relocate the interfering equipment l Increase the separation between the equipment and the possible source of the interference l Connect the equipment to an outlet on a circuit different from that to which the interfering equipment was connected Information regarding Electro Magnetic Compatibility EMC according to IEC60601 1 2 With the increa...

Страница 12: ...urer and d is the recommended separation distance in meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol Note1 At 80 MHz and 800 MHz the higher frequency range applies Note2 These guidelines may not...

Отзывы: