background image

I

f the unit gives an error code, immediately switch the unit off. See error table below for proper corrective action. If the error cannot be cleared, please contact your Talboys 

representative for repairs.  

T

roUblesHooTing

 (H

oTPlaTe

 / H

oTPlaTe

-s

Tirrers

)

The stirring units do not have LED displays. In the event of a malfunction, the unit will beep 10 (ten) times. To clear the error, turn the control knob to the off position and cycle 

the power by disconnecting the power cord from the source. Reconnect power cord and resume operations. If the error persists, turn the unit off and contact your Talboys 

representative for repair. 

11

Error

Cause of Error

How to Fix

E1

An “E1” error means the heater sensor is open or malfunctioned. 

This error cannot be fixed by the end user. Please contact your Talboys representative 

for repair.

E2

An “E2” error means the heater sensor shorted or malfunctioned.

This error cannot be fixed by the end user. Please contact your Talboys representative 

for repair.

E3

An “E3” error means there is either no motion on the motor or the motor is 

not working properly.

Reset the unit by rotating the knobs for speed and heat to the off positions until 

they click then turn them back on. If it still doesn’t work, please contact you Talboys 

representative for repair.

E6

An “E6” error means there is an internal electronics system error. 

This error cannot be fixed by the end user. Please contact your Talboys representative 

for repair.

E8

An “E8” error means the unit had a catastrophic over temperature condition 

or temperature runaway condition (temperature greater than 600°C) and 

therefore automatically shut down to prevent damage.

This error cannot be fixed by the end user. Please contact your Talboys representative 

for repair.

E9

An “E9” error means the heater failed. This might occur if heater temperature 

fails to rise when asked to, or there is a sudden drop in heater temperature 

for no apparent reason.

This error cannot be fixed by the end user. Please contact your Talboys representative 

for repair.

T

roUblesHooTing

 (s

Tirrrers

)

Содержание Advanced 10 X 10

Страница 1: ...INSTRUCTION MANUAL Advanced Hotplates Stirrers Hotplate Stirrers...

Страница 2: ...cturer warrants this product to be free from defects in material and workmanship when used under normal conditions For your reference make a note of the serial number date of purchase and supplier her...

Страница 3: ...el CLEANING ALUMINUM TOPS For simple dust and dirt clean the aluminum top by using a damp cloth with soap and water For more stubborn deposits try using a flat edge wooden spatula to scrape off as muc...

Страница 4: ...lammable materials Keep the power cord away from the heater plate DO NOT operate the unit at high temperatures without a vessel sample on the top plate WARNING Units are NOT explosion proof Use cautio...

Страница 5: ...eut tre d sactiv e si le dispositif est utilis avec des accessoires non fournis ni recommand s par le fabricant ou s il est utilis de mani re non stipul epar le fabricant Utilisez toujours le disposit...

Страница 6: ...minum ambient 5 C to 400 C Temperature stability ceramic 3 aluminum 2 Speed range 60 to 1600rpm Speed stability 2 Capacity 600mL Controls see diagram Ship weight 6 2lbs 2 8kg Below 100 C 2 C Environme...

Страница 7: ...r 120V 8 3 amps 1000 watts 230V 4 6 amps 1050 watts Fuses HP HPS 5mm x 20mm 10 amp slow blow Stirrer 5mm x 20mm 5 quick acting Temperature range ceramic ambient 5 C to 500 C aluminum ambient 5 C to 40...

Страница 8: ...atts Hotplate Stirrer 120V 11 2 amps 1345 watts 230V 7 0 amps 1600 watts Fuses 120V HP HPS 1 1 4 x 1 4 in 15 amp quick acting internal 230V HP HPS 5mm x 20mm 16 amp slow blow Stirrer 5mm x 20mm 5 quic...

Страница 9: ...late is too hot to touch The caution hot top indicator light will illuminate when the heat is turned on and remain on until the top plate cools down Recall last temperature button Advanced Series Hotp...

Страница 10: ...on e Selecting the SEt option saves the Single Point Calibration adjustment for that temperature set point and allows you to exit this SPC mode see Section c When the Set option is selected SEt will b...

Страница 11: ...s and limits allowed by the particular unit 10 Stirring Operating Instructions The micro processor controlled ramping feature slowly increases speed until the set point is reached This feature helps t...

Страница 12: ...d user Please contact your Talboys representative for repair E3 An E3 error means there is either no motion on the motor or the motor is not working properly Reset the unit by rotating the knobs for s...

Страница 13: ...e Advanced Heating Units only switch 386104 00 button 386101 00 5 Support stand thumb knob 186208 00 6 Housing 286613 00 7 Heat Shield 286607 00 8 Magnet assembly Stirrer Hotplate Stirrer 286616 00 9...

Страница 14: ...Heating Units only switch 386104 00 button 386101 00 5 Support stand thumb knob 186208 00 6 Housing 286614 00 7 Heat Shield 286608 00 8 Magnet assembly Stirrer Hotplate Stirrer 286616 00 9 IEC module...

Страница 15: ...tirrer Hotplate Stirrer 286616 00 9 Motor 286627 00 10 Main PCB Hotplate 386660 00 11 Pot PCB Assy 386661 00 12 Display PCB Hotplate Hotplate Stirrer 386663 00 Stirrer 386662 00 13 Bottom Plate 286612...

Страница 16: ...g Units only switch 386100 00 button 386101 00 5 Support stand thumb knob 186208 00 6 Housing 286615 00 7 Heat Shield 286609 00 8 Magnet assembly Stirrer Hotplate Stirrer 286616 00 9 IEC module 386681...

Отзывы: