The Orbital/Microplate Shakers have been designed for the speed and time functions 
to work independently of one another. The speed can be reset without resetting 
the timer and the timer can be stopped and started without interrupting the shaking 
function.
1.  getting ready:
  a.   Plug the power cord into a properly ground outlet. The standby indicator light 

will illuminate, verifying power to the unit.

  b.  Press the standby button to move the unit from standby mode. The standby 

indicator light will turn off and the speed and time displays will illuminate, 
displaying previously used settings.

2.  Setting speed:
  a.   Press the up/down arrows below the speed display until you reach the desired 

speed. When you release the button, the display will blink off and then on 
indicating the new set speed has been accepted.

  b.   Press the on/off button to start the shaking function. The indicator light below 

the speed display will illuminate to indicate the shaking function is in use and 
remain lit until shaking has ceased. The microprocessor controlled ramping feature 
slowly increases speed until the set-point is reached which helps to avoid 
splashing, and provides excellent low end control.

  c.   Speed adjustments can be made without interrupting shaking by using the up/

down arrows below the speed display. After the change has been made and 
you release the button, the display will blink off and then on indicating the new 
set speed has been accepted.

  d.   To stop the shaking function, press the on/off button below the speed display. 

The speed indicator light will turn off.

3.  Setting time to zero (0:00) and continuous mode: Accumulated time.
  a.   Press and hold the on/off button below the time display. After three (3) seconds 

the display will indicate the previous set time.

  b.   Simultaneously press both the up and down arrows, the display will indicate 

zero (0:00). The unit time is now set to zero (0:00) minutes. Alternately, you can 
use the up/down arrows to get to zero (0:00).

  c.  Press the on/off button below the time display. The display will indicate 

accumulated time. The up/down arrows will become inactive. To stop timer, 
press the on/off button again. iMPortAnt: This will not interrupt the shaking 
function. Press the on/off button below the speed display to interrupt the 
shaking function.

  d.  To reset, press and hold the on/off button below the time display. After three (3) 

seconds the display will indicate the previous set time, which was zero (0:00).

4.  Setting timed mode: Programmed time.
  a.   Press the up/down arrows below the time display until you reach the desired 

time.

  b.   Start this function by pressing the on/off button below the time display. The unit 

will run for the selected time, the up/down arrows will become inactive while the 
timer is running. The unit will stop shaking when the time display reaches zero 
(0:00). Four (4) audible beeps will indicate the count down function is complete. 
The time display will default back to the set time. To repeat for the same time, 
simply press the on/off button again.

  c.   To interrupt an automatic timing cycle before it is completed, press the on/off 

button below the time display. The display will flash off and on to indicate the 
time function is on “hold”. iMPortAnt: This will not interrupt the shaking 
function. Press the on/off button below the speed display to interrupt the 
shaking function. Restart the timer by pressing the on/off button below the 
time display. Unit will continue counting down to zero (0:00). When the display 
reaches zero (0:00), you will hear the four (4) audible beeps that indicate the 
count down function is complete and shaking function will cease.

a

DvanCeD

 o

rbital

/M

iCroPlate

 s

haker

 o

Perating

 i

nstruCtions

9

Содержание 1000-3

Страница 1: ...Instruction Manual Advanced 1000 3 Orbital Shaker Advanced 1000MP Microplate Shaker...

Страница 2: ...___ Supplier ____________________________________________________________ Table of Contents Package Contents 1 Warranty 1 Installation 2 Maintenance Servicing 2 Equipment Disposal 2 Environmental Cond...

Страница 3: ...gent or solvent on the front panel which is abrasive or harmful to plastics nor one which is flammable Always ensure the power is disconnected from the unit prior to any cleaning If the unit ever requ...

Страница 4: ...immerse the unit for cleaning Do not operate the unit if it shows signs of electrical or mechanical damage Earth Ground Protective Conductor Terminal Alternating Current Henry Troemner LLC hereby dec...

Страница 5: ...again be in standby mode B Speed display Displays the speed of the shaker C Up down arrows for set point control On off button starts stops shaking function D The speed indicator light will be illumin...

Страница 6: ...mm x 20mm 5 amp quick acting Speed range 100 to 1200rpm Speed accuracy 2 Timer 1 second to 9999 minutes increased in 1 second increments Orbit 0 125 3mm Maximum weight capacity 8lbs 3 6kg up to 1000rp...

Страница 7: ...for use with flask media bottle clamps or test tube racks See below for tray configurations Advanced 1000 3 Orbital Shaker Tray Configurations 35 10mL Erlenmeyer Flask Clamps 20 25mL Erlenmeyer Flask...

Страница 8: ...ts Fuses 5mm x 20mm 5 amp quick acting Speed range 100 to 1200rpm Speed accuracy 2 Timer 1 second to 9999 minutes increased in 1 second increments Orbit 0 125 3mm Capacity 4 microplates or 2 micro tub...

Страница 9: ...1 Place two 2 microplates or deep well blocks diagonally on the tray or place four 4 microplates or deep well blocks on the tray The plates blocks do not have to be filled 2 Place the corner of the p...

Страница 10: ...dicate the previous set time b Simultaneously press both the up and down arrows the display will indicate zero 0 00 The unit time is now set to zero 0 00 minutes Alternately you can use the up down ar...

Страница 11: ...exceeded loose foot suction cup Press the standby button to clear this error Be sure the load is within the maximum load capacity before restarting the unit If the E04 error persists switch the unit o...

Страница 12: ...385703 00 2 Tray 580050 00 3 Display board 380780 00 4 Belt 580019 00 5 Motor 380712 00 6 Motor transition board 380415 00 7 Connection cable 380720 00 8 IEC power entry module 386122 00 9 Feet sucti...

Страница 13: ...5702 00 2 Tray assembly 880761 00 3 Display board 380780 00 4 Belt 580019 00 5 Motor 380712 00 6 Motor transition board 380415 00 7 Connection cable 380720 00 8 IEC power entry module 386122 00 9 Feet...

Страница 14: ...2 980199 1 5 to 2mL Micro Tube Rack 70 2 980191 Installation of Flask Clamps Orbital Shaker 1 Unit should be in standby mode 2 Remove non skid rubber mat 3 Line up the hole s in the flask clamp to th...

Страница 15: ...Universal Harness Platform Part Description Microplate Capacity Capacity Number Universal Harness 2 or 4 1 980192 Installation of 2 Bar Adjustable Platform Orbital Shaker 1 Unit should be in standby...

Страница 16: ...201 Wolf Drive PO Box 87 Thorofare NJ 08086 Phone 1 888 TALBOYS Fax 856 686 1601 E mail troemner troemner com www troemner com 715082 00 REV 6 5 15 Manufactured by...

Отзывы: