Dice
Touch
Quick Guide -12
TAKARA BIO INC.
7.
Optional setting
Touch a 'Configuration' from Home screen.
1.
System Information
2.
Machine Settings
3.
Time & Date
4.
Thermal Calibration
5.
User Accounts
6.
Software Upgrade
7.
Touchpanel Calibration
1. System Information
Confirm the situation of the unit.
・
Serial Number
・
Controller Software Version
・
Firmware Version
Allows changes to be made to he instrument
2.
Machine Settings
settimgs.
Use predicted liquid temperature for hold times
:
Using predicted temperature effectively increase
the hold times and total run time. [Default
:
✔
]
(Based on a liquid volume is 10
~
50μl)
Accelerated Thermal Equilibration
:
[Defalt
:
None]
Selecting
this option will use block overshoot and undershoot control to accelerate thermal equilibration.
(Optimized for reaction volumes 20
~
30μl)
Auto Resume Run After Power Outage
:
[Default:
✔
]
Cause the instrument to automatically resume a run from the point where a power inrerruption occurs.
Rate limit heating cooling rates
:
[Defalt
:
None]
Set rate limit for a run. This has a limit of 0.1
~
2,5[Max]
℃
. (Only available in runs not manual mode)
Use Top Heater
:
[Default:
✔
]
Cause the top heater to be used during a run if it is checked. Entry box for top heater temperature.
Temperature between 30
℃~
115
℃ (
Default 105
℃)
. 'Turn on Top heater NOW' button for manually
enabling the top heater to preheat it if required. (Enable top heater when set temperature >30
℃
)
Used to set the time for this instruments
3. Time & Date
internal clock.
Set the day and time using a ca
l
ender tha
t
appear to click small arrow near the year and month.
Audible alarm on run completion
:
[Default:
Alert an operator that run has completed.
①
②
③
④
⑤
⑥
⑦