![TAITEC SR-2DW Скачать руководство пользователя страница 33](http://html.mh-extra.com/html/taitec/sr-2dw/sr-2dw_operation-manual_814207033.webp)
32
Trouble Shooting
If you find any trouble on the unit, check it referring the list below.
Problem
Check
When not improved.
Can not power on
Plugged in the wall outlet?
Possible to be blown the
fuse. Replace it with a new
one. May be caused by short
circuit when the fuse blown
again.
Voltage on the wall outlet is normal?
Power switch is “ON”?
Does not shake
Shake switch is “ON”?
Possible to be caused by
a motor, PCB and a broken
belt.
Speed volume is lowest?
“TIMER MODE” is ON?
Does not excess the max load?
Continue
shaking
displaying ”Errn”
Use power supply from 200V/3 phase down to
100V?
Check the wall outlet and
change the power supply if
needed.
If not caused by power
supply, possible to be
caused by a broken motor or
PCB.
Timer does not work
or display “ErrJ”
Shaker switch is “ON”?
PCB is broken.
Set timer is “0:00”?
Display “Err0”
PCB has a problem.
Big shaking noise or
abnormal noise.
Bolts on the shaking plate are loosen?
Possible to be deteriorated
parts of shaking inside.
Flasks are attached on the holder firmly?
A part of the shaking plate contacts
somewhere?
Unstable floor?
Main body moves.
Overweigh the max load?
Possible to be deteriorated
rubber stands.
Unstable floor?
The floor is sandy, dusty or slippery?
※※If it does not improve after checking the list above or if you have any questions other
than above, please contact your distributor.