4. Replace and twist the cap to close
the lock as shown in the picture until you
hear the click.
When the lock is properly closed lock
indicator is in the extreme right position.
Hold your fingers on the
body NOT the cap.
1
2
3
INSTRUCTIONS FOR USING
Setting up the Lancing Device
Alternative Site Testing (AST)
Cleaning
x
The Lancing Device allows you to collect your capillary blood from fingertip (and
alternative sites) easily and virtually pain free.
2. Insert a lancet into the lancet holder
and push down firmly until it is fully
secured.
3. Twist the protective disk off the lancet.
7. Lancing the finger
Press the lancing device’s tip firmly
against the lower side of your fingertip.
Press the release button to prick your
finger, then a click indicates that the
puncture is complete.
8. Remove the device from the puncture
site. After penetration, remove the first
drop of blood with a clean cotton swab.
Gently squeeze the punctured area to
obtain another drop of blood. Be careful
NOT to smear the blood sample. Apply
the blood to the test strip.
10. Slide the ejector forward to remove
the used lancet. Dispose of the lancet
according to your local regulations.
?
For collecting blood samples from
sites other than the fingertip, use the
clear cap.
Testing on sites other than the fingertip is referred
Please read the
Alternative Site Testing
section in your
you perform AST.
Substitute the lancing device cap with the clear cap
alternative site testing. Pull the cocking
When lancing the forearm, upper arm, hand, thigh, or calf, avoid lancing the areas
with obvious veins in order to prevent excess bleeding.
9. Remove the lancet by twisting the
cap off first. Safely dispose of the used
lancet by placing the protective disk on a
hard surface, and pushing the exposed
tip into the disk.
5. Holding the cap select on the depth of
penetration by turning the adjustable tip
in either direction so that the arrow on the
cap points to the desired depth.
6. Pull the cocking control back until it
clicks. You will see a color change inside
the release button when it is ready.
?
If it does not click, the device may
have been cocked when the lancet
Referent
was inserted.
The Lancing Device offers you 6 depths of skin penetration:
Symbol
The longer the length of the indicator line, the greater the depth of penetration.
Shallow
Deep
1
2
1
2
1. Twist and pull off the cap of the
lancing device.
To reduce the chance of infection:
? Never share a lancet or the lancing device.
? Always use a new, sterile lancet. Lancets are for single use only.
? Avoid getting hand lotion, oils, dirt or debris in or on the lancets and the lancing
device.
? The used lancet may potentially be biohazardous. Discard it according to your
WARNING
local regulations.
Before collecting a drop of blood:
? Wash and dry your hands before you start.
? Select the puncture site at the fingertip.
? Rub the puncture site for about 20 seconds before penetration.
NOTE
? Clean the site using cotton swab moistened with 70% alcohol and let it air dry.
Community
Sterilized using irradiation
Do not use if package is damaged
STERILE R
Ver 1.0 2014-12
318-5010200-xxx
Cap
Cocking
Control
Release
Button
Lancet Holder
Ejector
Arrow
Depth Indicator
Adjustable Tip
A Clear Cap
(For AST use only)
Lock Indicator
Lock
The lancet is disposable and is intended
single use only. The lancing device is reusable.
To clean the lancing device, use a mild soap and
water to wipe clean the lancing device.
DO NOT rinse the device under water. Disinfect
the cap by placing it in 70% alcohol for 10 minutes
and then allow it to air dry. Do NOT place the device
in a dishwasher or use detergents.
for
Soft or thin skin
Average skin
Thick or calloused skin
Marketed and Distributed by:
Rapid Check Diagnostics & Wellness Ltd.
11, Balogun Street, Off Obafemi Awolowo Way,
Ikeja, Lagos, Nigeria. P.O Box 499 Ikeja
Tel: 234-1- 892 9914; +234- 803 411 2811; +234- 807 189 1575
Email: [email protected], [email protected]
website: www.rapidcheckng.com
KIPTRACK LANCING DEVICE
to as alternative site testing (AST).
Owner’s Manual before
for
control back until it clicks.
Symbol Information
Do not reuse
Consult instructions before use
Use by / Expiry date
Symbol Referent
Batch code
Manufacturer
Caution, consult accompanying documents
Authorized representative in the European