4 Controls and Their Operation
4.5 Analog-Out Settings Menu
To Change Analog-Out Settings
1 From the menu screen, press the ANALOG OUT SETTINGS button.
2 The analog output settings menu will be displayed. Pressing the desired buttons allows you to set/change
the 4-20 mA.
3 Use the keypad to enter the values and press ENTER to save.
4 Press the MENU button to return to the menu screen. Press the READINGS button to return to the main screen.
1
2
3
The 4-20mA can be simulated to verify proper output
1 Press the O2 4mA button.
2 Adjust your multimeter to mA mode.
3 The value you enter on the controller should match the display on
the multimeter.
4 Press the MENU button to return to the menu screen. Press the
READINGS button to return to the main screen.