![Tabor Electronics Preliminary Lucid-X Series Скачать руководство пользователя страница 31](http://html1.mh-extra.com/html/tabor-electronics/preliminary-lucid-x-series/preliminary-lucid-x-series_user-manual_4369310031.webp)
Lucid-X Series RF Analog Signal
Generator PXIe Module User Manual
Rev. 1.0
Confidential | 31
4.5
Modulation
Table 4.5 Modulation Specifications
Modulation
Frequency Modulation
Maximum Deviation
10 MHz
Resolution
0.1 % or 1 Hz (the greater)
Modulation Rate
1 MHz
Resolution
1 Hz
Amplitude Modulation
AM Depth
Type
Linear
Maximum Settable
100%
Resolution
0.1 % of depth
Modulation Rate
DC to 100 kHz
Phase Modulation
Peak Deviation
360 deg
Modulation Rate
DC to 100 kHz
Pulse Modulation (PLS Option)
On/off Ratio
70dB
Rise/fall Time (10%-90%)
15 ns (typ.)
Resolution
10 ns
Minimum Width
30 ns
Repetition Frequency
DC to 10 MHz
Pattern Modulation (PAT Option)
Number of Steps
1 to 2,048
Step Repetitions
1 to 65,535
ON/Off Time
20 ns to 20 days
Sweep
Range
Same as frequency range
Modes
Frequency and amplitude step, list
Dwell Time
100
μs to 1,000 s
Resolution
1 μs