Discovery DSC™ Getting Started Guide
Page 23
**While performance during heating is not affected, during cooling the cooler is able to maintain cooling rate linear-
ity only below 325°C.
***Air cool can also be used when selected to operate when the cell cover is opened while operating with an active
cooling accessory to prevent moisture from entering the cell/cooler. In this case, the air cool inlet will be connected to
a dry nitrogen source.
CAUTION: When using the coated lid, the upper temperature of the instrument is limited to 400°C.
Operation above this temperature can damage the coated lid. If temperatures higher than 400°C are
required, make sure to install the standard silver lid prior to starting the experiment. Recalibration
may be required when switching between the two lids.
MISE EN GARDE: Lors de l'utilisation du couvercle revêtu, la température supérieure de l'instru-
ment est limitée à 400°C. Toute utilisation à une température supérieure à celle-ci peut endommager
le couvercle revêtu. S'il est nécessaire de travailler à des températures supérieures à 400°C, assurez-
vous d'installer un couvercle en argent standard avant de commencer l'expérience. Le réétalonnage
peut s'avérer nécessaire en cas de changement de couvercle.
CAUTION: Holding the Discovery DSC™ cell at high temperatures for extended periods of time
may cause sticking of the silver lid. If this occurs, please contact TA Instruments for assistance in
removing the lid.
MISE EN GARDE: Le maintien de la cellule Discovery à de hautes températures pendant de longues
périodes de temps peut engluer le couvercle en argent. Si cette situation se produit, contactez TA
Instruments pour obtenir de l'aide pour le retrait du couvercle.