Nano ITC Getting Started Guide
Page 35
CAUTION: Always use proper protective equipment when working with samples and cleaning
fluids.
CAUTION: Thoroughly rinse all areas that come into contact with corrosive chemicals.
CAUTION: All rinsing or flushing operations should be done with the cleaning devices described
beginning on page 33. Caustic solutions should be loaded and removed with the appropriate
syringe or micropipette.
Soapy Water Flush
1
Exit from the ITCRun software. (This must be done whenever you pass large volumes of solutions
through the Nano ITC.)
2
Flush or aspirate 100 mL of dilute aqueous detergent solution (
e.g
. SDS) through the cells. Make sure
to use a detergent that does not leave any residue.
Final Flush
Flush or aspirate 1 L of deionized water through each cell.