![T Scale RWP Скачать руководство пользователя страница 23](http://html1.mh-extra.com/html/t-scale/rwp/rwp_technical-manual_3069356023.webp)
RWP Weighing Indicator Technical Manual(v1.16)
- 23 -
Load the first calibration mass weight on the platform (mass weight should
be1/3 of the max capacity) and wait few seconds for display stability.
Then press
key , display will be shown
Load the second calibration mass weight on the platform (mass weight
should be2/3 of the max capacity) and wait few seconds for display
stability.
Then press
key ,display will be shown
Load the third calibration mass weight on the platform (mass weight should
be3/3 of the max capacity) and wait few seconds for display stability.
Then press
key ,display will be shown
After the calibration the display will start a self test. Remove the load from
platform during the test. Display will come to weighing mode automatically.
Creep Calibration
Enter the function by pressing
, display will be shown
Make sure there are no loads on the platform and wait few seconds for
stable indicator on.
Enter the function by pressing
, display will be shown
Load the calibration mass weight on the platform (mass weight should be
the max capacity) and wait few seconds for display stability.
Then press
,display will be shown
Remove the weight ,display will be shown
After the calibration the display will start a self test. And come to weighing mode
automatically
If display will be shown any error or incorrect value, repeat the procedure again.
Load 2
Load 3
pass
unLoad
Load F
Creep xx
Creep OK