AW20 Series Weighing Scale Operation Manual
- 43 -
6. CALIBRATION
6.1 Normal Calibration
▪
Press
key to enter setup menu.
▪
Using numeric keys
、
、
、
to enter calibration.
6. Service> 3.calibration parameters >
4.SCALE1 > 2.Calibration> 1.Normal Calibration
▪
Before access to the parameter,it need to press calibration switch,locating back side of indicator in
approval version.
Steps
Remark
1 Press
Key to enter.
The display will be show
“
Please empty
the scale
”
2 Remove goods from the platter.
3 Press
key to confirm.
The display will be show
“
Enter
calibration weight by using numeric
keys
”
.