MFT-PRO
Quick reference guide
EN – 18
Test on earth leakage relay RCD (with accessory RCDX10)
Residual current protection type (RCD):
AC ( ), A (
), B(
) – General (G), Selective (S) and Delayed ( )
Voltage range Phase-Earth, Phase-Neutral:
100V
265V RCD type AC and A, 190V
265V RCD type B
Rated tripping currents (I
N): 0.3A
10A
Frequency: 50/60Hz
5%
Earth leakage relay RCD tripping current ( ) – (for RCD General only)
RCD type
I
N Range
I
N
[mA]
Resolution [mA]
Accuracy
AC, A, B
300mA
I
N
10A (0.3
1.1) I
N
0.1I
N
- 0%, +5%I
N
Duration of earth leakage replay RCD tripping time test – TT/TN systems
x 1/2
x 1
x 2
x 5
AUTO
\ G
S G S G S G S G S G S
0.3A
1.0A
AC 999 999 999 999 999 999 200 250 50 150
310
A 999 999 999 999 999 999 200 250 50 150
310
B 999 999 999 999 999 999
310
1.1A
3.0A
AC 999 999 999 999 999 999 200 250 50 150
310
A 999 999 999 999 999 999 200 250 50 150
310
B 999 999 999 999 999 999
3.1A
6.5A
AC 999 999 999 999 999 999 200 250 50 150
310
A 999 999 999 999 999 999 200 250 50 150
310
B 999 999 999 999 999 999
6.6A
10.0A
AC 999 999 999 999 999 999 200 250
A 999 999 999 999 999 999
B
Table with duration of tripping time measurement [ms] - Resolution: 1ms, Accuracy:
(2.0
rdg + 2dgt)
Duration of earth leakage replay RCD tripping time test – IT systems
x 1/2
x 1
x 2
x 5
AUTO
\ G
S G S G S G S G S G S
0.3A
3.0A
AC 999 999 999 999 999 999 200 250 50 150
310
A
B
3.1A
6.5A
AC 999 999 999 999 999 999 200 250 50 150
310
A
B
6.6A
10.0A
AC 999 999 999 999 999 999 200 250
A
B
Table with duration of tripping time measurement [ms] - Resolution: 1ms, Accuracy:
(2.0
rdg + 2dgt)
NoTrip – Global earth resistance with no residual current protection tripping
Voltage range Phase-Earth, Phase-Neutral:
100V
265V
Frequency: 50/60Hz
5%
Global earth resistance in systems with Neutral
Range [
] Resolution
[
]
Accuracy
0.01
9.99
0.01
-
(5
rdg + 0.1
)
10.0
199.9
0.1
-
(5
rdg + 1
)
200
1999
1
-
(5
rdg + 3
)
Maximum current:
<15mA
Ut LIM (UL):
25V or 50V
Global earth resistance in systems without Neutral
Range [
] Resolution
[
]
Accuracy
1
1999
1
-
(5
rdg + 3
)
Maximum current:
< ½ I
N
set
Ut LIM (UL):
25V or 50V