SZBJ BM852A+ Скачать руководство пользователя страница 2

  Cautions

When the temperature detector head not put into the meter, the displaying 
value is meaningless, attached with the meter, the limited temperature of the 
conjunction thermoelectric couple , which is the K type WRNM-010, bareness 
style, is 250

( may reach 300

 in short time)

 

 
4.10 Measurement of frequency   
Turn  the  range  switch  into  the  range  2kHz,  put  the  black  pen  into  the  jack 
“COM”,  the  red  into  the  jack“V/

”,  in  parallel  connect  the  two  pens  into  the 

measured circuit, get the reading. 
4.11 Measurement of phase sequence 
Turn  the  range  switch  into  the  range  AC600V,  put  the  black  pen  into  the 
jack“COM”,  the  red  into  the  jack  “V/

”,  take  a  press  on  the  button  for 

measurement  of  phase  sequence,  LCD  shall  display  symbol“

  ”

for 

measuring phase sequence, after a measure to phase sequence, if measuring 
again,  it  is  necessary  to  press  the  button  for  phase  sequence  to  cancel  the 
function on measurement of phase sequence at first, then restart the function 
on measurement of phase sequence, the meter sets its phase sequence as   
A          B          C. 
4.11.1 Measurement of three-phase four-wire system: 
(1)

 

Confirm  phase  A:  connect  the  black  pen  into  the  zero  wire  or  the 
ground  wire,  the  red  pen  into  any  hot  wire,  LCD  shall  show  the 
symbol A,  here the  phase connected  with  the  red  pen  is  defined  as 
phase A. 

(2)

 

Find phase B or phase C: connect the black pen into the zero wire or 
the ground wire, the red into any hot wire of the other hot wires, when 
LCD  shows the  symbol  B, the  phase  connected  with  the  red  pen  is 
defined  as  phase  B

when  LCD  shows  the  symbol  C,  the  phase 

connected with the red pen is defined as phase C 

(3)

 

During measuring, the meter may show the value of the line voltage 
of the measured phase: 

(a)  When  the  measured  line  voltage  is  less  than  30%  of  the  other 

phases, then the phase is lost; 

(b) When either phase B or phase C cannot be found, then either phase 
B or phase C is lost. 

4.11.2 Measurement of three-phase three-wire system

 

(1)

 

  Confirm phase A: connect the black pen into any wire, the red pen into 
any  wire  of  the  other,  LCD  shall  show  the  symbol  A,  here  the  phase 
connected with the black pen is defined as phase A. 

(2)

 

  Find phase B or phase C: 
(a)

 

Keep the black on the original phase, connect the red into the third 

phase, when LCD showing B, here the phase connected with the red 
is phase B, may judge the phase connected with the red in step (1) is 
phase C. 

(b)

 

When  LCD  showing  C,  here  the  phase  connected  with  the  red  is 

phase  C,  may  judge  the  phase connected  with the  red  in step  (1)  is 
phase B. 

(3)

 

  During  measuring  to  step  (1)

(2),  the  meter  shall  show  line  voltage

380V

. If the voltage of a phase is less than 30% of the others, then 

the phase of the lower voltage is lost, after phase A is confirmed, either 
phase B or phase C cannot be found, then either phase B or phase C is 
lost. 
 

Cautions 

please reset before each measurement

quit the function 

on  the  measurement  of  phase  sequence,  then restart  the  function

measuring period may not more than 10 seconds. 

5. Technological Indices   

Accuracy guarantee: 1 year

Temperature of keeping accuracy:23

±5

   

Relative humidity:<75 %

 

Function 

Range 

Resolution 

Accuracy 

20A 

10mA 

ACA 

600A 

1A 

 

1.9%

ACV 

600V 

1V 

1.2%

5d 

20V 

0.01V 

DCV 

600V 

1V 

0.8%

2d 

Resistor 

2k

 

1

 

1.0%

2d 

FREQ 

2kHz 

1Hz 

2.0%

5d 

ACA  frequency  range 

50~60Hz(  sine  wave)

ACV  frequency  range

50~100Hz(sine wave)

 

 

Function   

Range 

Resolution 

Accuracy 

-30~400

 

1

 

1.2%

4d 

Temperature 

400~1000

 

1

 

1.9%

15d 

 

Diode 

 

Show the value of the forward voltage. Test conditions: 

forward DC current is about 1mA

backward DC voltage 

is about3V

 

Connection & 

disconnection 

 

When Conductive resistance is less than about 50

, the 

buzzer in the meter sounds, test conditions

open voltage 

is about 3V

 

Phase 

sequence 

May  measure  the  phase  sequence  of  AC  when  phase 
tolerance is under ±25 

Resistor,Diode;Buzzer,FREQ    Overload Protection :250V   
 
6. Maintenance of the Meter 
 

   

Cautions

Before open the covers of the meter, or the batteries, shall 

switch  off  power  and  disconnect  the  pens  and  any  input  signal  to  prevent 
danger of shock. 
 
6.1  when  the  meter  shows  symbol“

”,  shall  change  battery.  Open  the 

battery cover, change a fresh battery with 9V ( Automatic range meter

Two 

new batteries with AAA1.5V ), to make sure the meter normal work. 
6.2 Keep the meter and the pens clean, dry and non-damage, may clean the 
meter surface  with clean clothes  or  eradicator,  abrasive  or  organic  solvent  is 
banned. 
6.3 Avoid mechanical damage, shake, impact, keep the meter away from high 
temperature and strong magnet field. 
6.4 The meter shall be calibrated once a year. 
 
7. Accessories   
 
7.1 A pair of pen 
7.2 A manual   
7.3    K type detector head for temperature 