31
MAXFRAME AUTOSTRUT™ System • Physician User Manual
Start the treatment.
The orange LED in the control system will turn off. Treat-
ment plan PDF contains private patient information and
will automatically be deleted from the laptop computer.
Disconnect USB cable.
After 30 seconds, each strut will lengthen by 0.1 mm.
Then, after another 2 minutes, each strut will shorten by
0.1 mm. This is done to verify the system works properly.
Replanning
When initiating a replan, if desired, surgeon can move
the struts to the nearest integer value as defined in the
MAXFRAME Software replan by utilizing manual soft-
ware mode or manual adjustment knob.