3.5 mm LCP
®
Extra-articular Distal Humerus Plate Surgical Technique DePuy Synthes 9
Preliminary Fixation
4
Preliminary fixation and compression
Instruments
310.25
2.5 mm Drill Bit, quick coupling
311.43
Handle, with quick coupling
314.02
Small Hexagonal Screwdriver with Holding
Sleeve
314.03
Small Hexagonal Screwdriver Shaft
323.36
3.5 mm Universal Drill Guide
399.121*
Bone Holding Forceps, 14 mm, serrated jaw,
soft ratchet
Alternative instrument
314.116 StarDrive
TM
Screwdriver Shaft, T15
After reducing the fracture, apply the plate and insert a
nonlocking screw through the center of the DCU portion
of an elongated Combi hole proximal to the fracture.
Use the 2.5 mm drill bit through the 3.5 mm universal drill
guide to predrill the bone. For the neutral position, press the
drill guide down in the nonthreaded hole. Use the depth
gauge to determine screw length.
Select and insert an appropriate length 3.5 mm cortex screw.
Do not completely tighten the screw. Make any final
adjustments to plate placement. Manually tighten the screw to
maintain plate placement and compress the plate to the bone.
* Also available