7
Trim, bend and cap wires
Leave 3 cm–4 cm in case retensioning becomes necessary.
8
Construct anterior half-frame
9
Connect ring to half-frame with ring-to-rod clamp
10
Reduce extra-articular component of fracture
Use ring and anterior rod as handles to manipulate fracture.
Tighten ring-to-rod clamp.
11
Add second rod
Connect ring-to-rod clamp and large combination clamp.
Add carbon fiber rod.
Note:
Operating instructions for the Wire Tensioner
(393.742) can be found in
Wire Tensioners for Hybrid Fixator
Synthes
10
Synthes