Synthes Air Powered Equipment

Cleaning

Before cleaning, remove all  
instruments, attachments and 
hose from the air drive unit. 

Do not immerse in water. 
Do not clean in automatic 
washer or ultrasonic  
cleaner. 

Use a neutral pH cleaning  
solution.

Clean and lubricate drive unit, 
attachments and air hose  
by hand immediately after 
each use.

General Maintenance

Reverse trigger
(both triggers)

Variable speed  
trigger

Attachment
release button

Air control valve

ComPact Air Drive II (511.701)

Air inlet

1

Manually clean air drive and 
attachments under running 
water. Use a mild cleaning  
solution (neutral pH) and a  
soft brush.

Do not allow water to  
enter the air inlet of the 
drive unit.

2

Clean cannulations of the air 
drive and attachments with 
the cleaning brush. Do not  
insert cleaning brush into the 
air inlet. 

The Quick Coupling for  
K-wires (511.791) must be 
fully open to its maximum 
diameter of 3.2 mm before 
inserting the brush.

3

Retract and clean the quick 
coupling hose end with a 
soft brush and mild cleaning 
solution. 

Do not allow water to  
enter the air inlets.

4

Clean hose with a mild  
cleaning solution and a soft 
brush or cloth.

Keep hose ends out of  
water.
Dry all instrumentation 
with a soft cloth.

Before cleaning, turn air control valve to the “off” position. Manipulate moving parts 
under running water to loosen and remove debris. Open and close chucks, and move 
quick couplings back and forth.

Lubricating

Lubricate the ComPact Air Drive II after every cleaning and before sterilization.
Use Synthes Autoclavable Oil (519.97).
Important: Never lubricate the radiolucent drive.

1

Before lubrication, open air 
control valve. Apply approx-
imately five drops of oil into 
the air intake of the drive 
unit. Lubricate triggers.

2

Connect the air drive to an air 
hose. Run the air drive for 20 
seconds in both directions.

3

Lubricate all moving parts of 
the attachments with one 
or two drops of oil. Operate 
the parts to distribute the oil. 
Wipe off excess oil.

4

Lubricate the quick coupling 
hose end with one or two 
drops of oil.

Forward 

(Lower trigger)

Reverse

(both triggers)

   