operation
2
3
Synrad
Firestar ti-Series
operator’s manual
Controls and indicators
ti-Series rear panel
Figure 2-2
ti-Series rear panel controls and indicators
1
DC Power Cables – re48 VDC from the DC power supply. Firestar ti-Series DC power cables
are manufactured from #10 AWG wire and measure 1 meter (42 inches) in length.
2
INT
(Remote Interlock) Indicator – illuminates green to indicate that a remote interlock circuit is
closed and that lasing may be enabled. The
INT
indicator is red and lasing is disabled if the interlock
input is open.
3
TMP
(Temperature) Indicator – illuminates green to indicate that laser temperature is within limits
and that lasing may be enabled. The
TMP
indicator is red and lasing is disabled if the laser’s tem-
perature rises above safe operating limits.
4
RDY
(Ready) Indicator – illuminates yellow when the laser is enabled, indicating that, after a five-
second delay, lasing will begin when a PWM Command signal is applied.
5
SHT
(Shutter) Indicator – illuminates blue to indicate that a
Shutter Open Request
signal is con-
nected to the
User I/O
port and lasing may be enabled. When a
Shutter Open Request
signal is
applied, there is a five-second delay until PWM inputs are recognized.
6
LASE
Indicator – illuminates red to indicate that the laser is actively lasing. The
LASE
indicator is
off when tickle pulses are being generated and illuminates red when PWM Command signal pulses
are long enough to produce laser output.
7
Keyswitch
(Keyswitch models only) – enables/disables operation of the laser. Firestar is enabled when
the
Keyswitch
is turned to the ON position. Turn the
Keyswitch
OFF to disable lasing.
8
USER I/O
Connector – provides a connection point for auxiliary output power as well as input and
output signals. Refer to
User I/O connections
in the Technical Reference chapter for pinouts and
signal descriptions.
9
WATER IN
Port (water-cooled models only) – labeled IN, this connection provides the cooling
water inlet to Firestar’s water-cooling system.
10
WATER OUT
Port (water-cooled models only) – labeled OUT, this connection provides the cooling
water outlet from Firestar’s water-cooling system.
ON
USER I/O
INT TMP RDY SHT LASE
(Fan-Cooled ti60 Shown)
1
2
4
5
3
6
8
7
9
10
OFF
Содержание Firestar TI Series
Страница 2: ......
Страница 4: ......
Страница 24: ...laser safety 12 Synrad Firestar ti Series operator s manual This page intentionally left blank...
Страница 40: ...getting started 116 Synrad Firestar ti Series operator s manual This page intentionally left blank...
Страница 104: ...maintenance troubleshooting 418 Synrad Firestar ti Series operator s manual This page intentionally left blank...
Страница 112: ...index i8 Synrad Firestar ti Series operator s manual This page intentionally left blank...