VelaShape III User Manual Maintenance
PB81499EN
75
Cleaning and Maintaining the VContour Applicator
The replaceable cover of the applicator should be replaced according to the schedule
printed on the cover's box label, or sooner if one of the following is observed:
●
After cleaning the replaceable cover and the applicator's pin connectors, an
Improve Vacuum Seal
message appears on the control panel (inadequate
coupling).
●
Electrodes appear damaged.
●
The cover's shape appears distorted.
Proceed as follows:
1.
After each treatment turn the system off and remove the applicator cover by
pulling it off.
2.
Remove the filter from the cover and wash or discard it.
3.
Wash the cover thoroughly using hot tap water (no more than 50°C/122°F)
and a mild disinfectant, alcohol-free soap until all residual VelaSpray Ease is
removed. Wipe and dry the cover with a soft cloth and store it in the patient's
kit. Re-use the cover only when completely dry.
4.
Allow the cover to dry overnight with the RF electrodes facing up.
5.
There are two filters per cover, one inserted and one spare; after each
treatment the filter should be removed, the spare filter should be inserted and
the used one washed, wiped with a soft cloth and allowed to dry in the patient
box until next the treatment.
6.
Clean the applicator's components
at the end of every day
as follows:
●
Refer to Figure 56: remove the applicator cover.
●
Clean the RF contacts (
A
), the ID microchip contact points (
B
) and the
vacuum inlet (
C
) with a clean, soft cloth.
●
Gently clean the IR LED’s aperture (
D
) using a gentle, alcohol-free optical
lens cleaning solution; be careful not to scratch it.
Figure 56: Applicator's Internal Components