VelaShape III User Manual Maintenance
PB81499EN
73
Note
The image in the figure is for illustration purposes only.
2.
If it becomes necessary to clean the applicator cover during treatment:
●
While the system is on, clean the rollers of the VSmooth
applicator while
pressing the trigger, with all modes at level
0
, while rolling over a mild
disinfectant-dampened tissue (that does not contain alcohol).
●
Turn the system off and remove the cover. Use hot tap water (no more
than 122°F/50°C). The cover should be flushed and washed thoroughly in
hot water and mild disinfectant soap or detergent that do not contain
alcohol, until all residual VelaSpray Ease is removed. Wipe and dry the
cover with a soft cloth. Reconnect the cover only when completely dry.
3.
Clean the VSmooth
applicator's interior components at the end of every day
as follows (see Figure 54):
●
Replace the two filters (see Figure 55) according to the schedule printed
on the cover's box. Never use a pair of filters for more than 10
accumulative hours.
●
Clean the RF contacts and ID contacts with a soft cloth.
●
Gently clean the IR lamp aperture using a gentle optical lens cleaning
solution, such as Fisherbrand
®
Optical Lens Cleaner or equivalent; be
careful not to scratch it.
●
Gently dry the IR lamp aperture with lint-free lens cleaning tissue.
●
Reconnect the clean and dry (or new) applicator's cover.
Figure 54: VSmooth Applicator's Internal Components