background image

IZ2/IZ2H Stimulator

Fast Facts

Status Light

solid red 

not  properly  connected  to 

base station or cannot sync

solid green 

connected to RZ and operating 

in current mode

solid green, slow red flash

 

 

connected to RZ and operating 

in voltage mode

Stim Lights

lit green 

 

indicated channel in use as a 

stimulus output

lit red 

 

indicated  channel  is  clipping 

(/- 10 V)

Battery Indicator Lights

VA  Positive Battery Pole
VB  Negative Battery Pole
VC  Logic Battery Level

Battery Status Lights

8 green

   

 

fully charged

1 green, 7 unlit 

low voltage

1 flashing red 

low voltage - charge now!

8 green flashing

  charging in progress

This  fast  facts  sheet  provides  basic  reference 
information for the IZ2/IZ2H Stimulator and related 
devices. See the System 3 Manual for more detailed 
information.

The  IZ2  Stimulator  is  available  with  32(IZ2-32),  64(IZ2-64),  or 
128(IZ2-128)  channels.  Power  for  stimulation  is  supplied  by  200 
and 400 Wh battery life (LZ48-200 and LZ48-400) Li-Poly battery 
packs, both capable of producing the same output voltage/current 
characteristics.  The  number  of  channels  needed  for  stimulation 
determines power requirements.  The IZ2H is a high current range 
version of the IZ2 and is available with sixteen stimulus channels.

Output

Stimulus Output Voltage 

+/- 12 V

IZ2 Stimulus Output C/- 300 µA up to 50 kOhm load
IZ2H Stimulus Output Current   +/- 3 mA up to 5 kOhm load 

Power

IZ2-128   should only be used with the LZ48-400
IZ2-64 

can be used with LZ48-200 or LZ48-400

IZ2-32  

can be used with LZ48-200 or LZ48-400

IZ2H-16  should only be used with the LZ48-400

Synapse Rig Configuration. 

 

The first time the IZ2 is used, 

it must be configured in the rig.  When the specialized DSPI 
card mounted in the RZ device is detected, an IZ2 will be 
added to the rig hardware tree. Click the check box to the 
left of the IZ2 icon to enable the device. In the options area, 
select the model and number of channels.

Synapse IZ2 Options. 

 

Click the IZ2 in the Processing Tree 

to display hardware option.  Select Current or Voltage mode 
and any other options as needed.

WITH

Отзывы: